100-214-294 REV. 05
21
4.3
Temperature
Table 4.8
Temperature
Item
Heater
The heater may cause some discoloration of the tank. This is normal
and will not affect the performance of the unit.
Solution
The fastest method to heat your ultrasonic bath is to fill with warm
solution, use heat, ultrasonics (which also adds heat), and a cover.
Over
Temperature
Protection
(CPXH only)
If Max Temperature of 75° C is reached, Ultrasonics will pause for a
period of time until the temperature in the bath falls back down to 69° C
(the max set point). Once the temperature falls back down to 69° C,
Ultrasonics will resume. This will apply for both continuous/infinity
mode, as well as timed mode. In timed mode, the timer will pause while
Ultrasonics is off, and will resume once Ultrasonics resumes.
Содержание Branson CPX1800
Страница 4: ...iv 100 214 294 REV 05 ...
Страница 6: ...vi 100 214 294 REV 05 ...
Страница 8: ...viii 100 214 294 REV 05 ...
Страница 17: ...100 214 294 REV 05 7 Chapter 2 Introduction 2 1 How Ultrasonics Works 8 2 2 Ultrasonic Baths 9 ...
Страница 21: ...100 214 294 REV 05 11 Chapter 3 Delivery and Handling 3 1 Unpacking Your Unit 12 ...
Страница 35: ...100 214 294 REV 05 27 Chapter 5 Installation and Setup 5 1 Installing Your Unit 28 ...
Страница 76: ...72 100 214 294 REV 05 ...