8 What must be done following failed operation
The TRA3000 generators have many different messages to assist the operator solving possible problems,
give information regarding incorrect operation of the TRANSIENT-generator, or to correct an incorrect
system configuration. Basically, three different messages can be differentiated:
•
Error message based on incorrect inputs
•
Error based on incorrect operation of the generator
•
Warning
messages
8.1.1 Error caused by incorrect inputs „Generator not ready for run“
Message Description
Emergency stop active
An emergency stop has been operated via the BNC outlet
„Emergency Stop“ on the rear of the TRA3000 F-S-D-V-C.
Variac input voltage fault
The variac could not be set to the correct value. Please check:
-Is voltage on EUT Power 1?
-variac bridges inserted on the rear panel?
-variac fuse o.k.?
Variac input frequency fault
The variac is only designed for 50/60Hz. Please check:
-Is frequency on EUT Power 1 out of range?
-variac bridges removed on the rear panel?
-variac fuse o.k.?
No Variac input
The variac has no power. Please check:
-Is voltage on EUT Power 1?
-variac bridges inserted on the rear panel?
-variac fuse o.k.?
No coupling path defined
In the „Main“ menu no coupling path for superimposing SURGE and
EFT onto the power line has been defined.
Repetition < 100ms
At EFT:
When synch mode = On, the Burst repetition must be greater than
100 ms.
No power input
Check power voltage on EUT input on rear of TRA3000 and the 16A
fuse.
Hardware power status fault
Press any of the front panel buttons. Repeat the test If there is no
change contact EMC PARTNER
Power must be 0V
Select in the Power menu “Main” instead of “0”.
Input power N> 50V and L-N <N
Interchange L and N
Содержание EXTTRA3000 E
Страница 8: ...TRANSIENT3000 8 162...
Страница 102: ...TRANSIENT3000 102 162...
Страница 106: ...TRANSIENT3000 106 162...
Страница 116: ...TRANSIENT 2000 116 162...
Страница 118: ...TRANSIENT 2000 118 162...
Страница 142: ...11BRemote Ports 142 162 All Generators of EMCP can be loaded and demonstrated...
Страница 148: ...12BAppendix and Corrections 148 162 13 1 6 VARIATION Specification IEC 61000 4 11 Ed 2...
Страница 150: ...12BAppendix and Corrections 150 162...
Страница 156: ...14BIndex 156 162...
Страница 158: ...14BIndex 158 162...
Страница 160: ...14BIndex 160 162...
Страница 162: ...14BIndex 162 162...