Do not use any excessive force when closing the sensor lid; while the tube should be fixed 

within the channel, it must not be crammed into it as this might affect the measurement 
values or damage the sensor.

• 

Try not to pull the tube into position as this might deform the tube and/or create cracks in 
the material which, in turn might impact the ultrasonic signal. 

• 

Ensure that the lid is fully closed to avoid fluctuating coupling and flow values. 

Please note:

• 

To avoid flow turbulences and associated measurement inaccuracies, it is necessary for the tube to be 

straight in the area near the BioProTT

TM

 Clamp-On SL.

 

Ideally, the tube should be straight for a distance of at least 15 x the inner diameter (ID) of the tube on 
either side of the sensor. 

For more information regarding sensor positioning and other possible measures to optimize and maintain 
the accuracy of the BioProTT

TM

 FlowMeasurement System, see our TechNote titled "

Optimizing Accuracy

". 

6.4  Use of the BioProTT

TM

 Clamp-On SL

For further details concerning the use of the BioProTT

TM

 Clamp-On SL and how to carry out flow 

measurements, please refer to the user manual of the respective BioProTT

TM

 evaluation device.  

Make sure to follow the steps described below and to heed the warnings included in them in 
order to ensure the highest possible accuracy of measurement values. 

In addition, always ensure that: 

• 

the medium temperature is the temperature the sensor was adjusted and calibrated for.

• 

a zero flow adjustment is carried out before each measurement

 and whenever possible 

without disturbing your process. 

Before starting the flow measurement:

1.

  Make sure that only tubes specified for the BioProTT

TM

 Clamp-On SL are used as this is the tube type the 

BioProTT

TM

 Clamp-On SL is adjusted and calibrated for. 

 

The tube that is used during the measurement must be the one indicated in the set calibration table.

 

The tube type(s) the sensor was adjusted and calibrated for can be found in the Calibration Information 
Sheet, which is part of every sensor order.  

2.

  Before each measurement, the cables, connectors and pins must be checked for breaks or damages 

as this could result in wrong measurements. Do not use the BioProTT

TM

 Clamp-On SL if you detect any 

damages. 

Figure 3: Inlet Section

BioProTT™ Clamp-On SL

Page  12

Copyright em-tec GmbH

D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 

DocuSign Envelope ID: BE88571C-E493-4DFB-AF9A-864EB3FC3D75

Содержание BioProTT Clamp-On SL

Страница 1: ...BioProTTTM Clamp On SL User Manual DocuSign Envelope ID BE88571C E493 4DFB AF9A 864EB3FC3D75...

Страница 2: ...c GmbH Finning All rights reserved Subject to Technical Changes Owing to our policy of continuous product development the illustrations and technical data contained in this document may differ slightl...

Страница 3: ...Maintenance and Service 9 6 Use of the BioProTTTM Clamp On SL 10 6 1 General Safety Information 10 6 2 Connecting the BioProTTTM Clamp On SL to the Evaluation Device 10 6 3 Attaching the BioProTTTM Cl...

Страница 4: ...itoring and control of the upstream feed monitoring and control of the downstream feed filtration fill and finish processes process evaluation upscaling processes research The SL at the end of the pro...

Страница 5: ...Clamp On SLs incorporate two pairs of transducers i e four piezo ceramics in total which are arranged in an X configuration around the tube containing the measured flow When sending ultrasonic signals...

Страница 6: ...aning Do not dispose of this device together with domestic waste The device as a whole as well as any parts must be disposed of in accordance with WEEE Directive and national legislation Manufacturer...

Страница 7: ...Clamp On SL is part of to ensure that the BioProTTTM Clamp On SL is used inspected and maintained in accordance with the user manual This also applies for all subsequent revisions or instructions from...

Страница 8: ...8 Troubleshooting as well as to the user manual of the evaluation device respectively 5 5 Electrical Safety and Electromagnetic Compatibility The BioProTTTM FlowMeasurement System i e the BioProTTTM C...

Страница 9: ...service for the BioProTTTM Clamp On SL may be carried out by em tec GmbH only If these instructions are not followed em tec GmbH shall accept no liability for the component and the warranty will be vo...

Страница 10: ...TTM FlowTrack SL To connect the BioProTTTM Clamp On SL to the BioProTTTM FlowTrack SL follow the steps listed below Carefully attach the round sensor plug to the sensor connection of the evalution dev...

Страница 11: ...t deformed damaged or dirty Ensure that the tube size and the tube material used corresponds with the size of the BioProTTTM Clamp On SL Air in the tube can lead to errors in the measurement or interr...

Страница 12: ...ease refer to the user manual of the respective BioProTTTM evaluation device Make sure to follow the steps described below and to heed the warnings included in them in order to ensure the highest poss...

Страница 13: ...ued or no flow value is displayed as the low coupling might impair the measurement It is important that any unwanted interruption of the process due to damage or kinking of the tubing is avoided at al...

Страница 14: ...ction of the BioProTTTM Clamp On SL Make sure that no detergents or disinfectants leak into the sensor plug For the cleaning of the device we recommend commonly used detergents and or the recommended...

Страница 15: ...ensor or to fix it in one position em tec offers optional mounting solutions The features differ slightly in regard to their design and handling depending on the sensor size Sensors with an OD of up t...

Страница 16: ...or torsial moment caused by the tube are too high the thread may tear the sensor base body may break or the sensor might be levered out If the mounting feature and thus sensor is not positioned corre...

Страница 17: ...mmends regular calibrations to be carried out Ideally the time between re calibrations should not exceed 24 months Other than the adjustment the calibration can be carried out by the customer on site...

Страница 18: ...nfluence the sensor or if the sensor or its cable came into contact with an electric cautery 2 Check the grounding of the valves and pumps or any other devices that might create electromagnetic distur...

Страница 19: ...nsor lid is closed properly 2 Close the lid If the problem persists contact our service department The tube size and or material is not the one indicated for the sensor 1 Check if the tube size and ma...

Страница 20: ...hannel If the problem persists contact our service department There might be electromagnetic fields influencing the sensor 1 Check if the sensor is damaged or not connected properly to the device 2 Ch...

Страница 21: ...sories must be disposed of in accordance with the applicable national provisions for electronic components In accordance with the requirements of EU Directive 2012 65 EC Waste Electrical and Electroni...

Страница 22: ...to OD 7 8 2 of reading 80 ml min up to OD 1 1 8 2 of reading 200 ml min up to OD 1 5 8 2 of reading 400 ml min The specified accuracies were determined under the following conditions Medium Water Med...

Страница 23: ...9 16 13433 PCT 3 8 x 1 8 0 017 2 5 0 100 10 3 8 1 8 5 8 13422 PCT 1 2 x 3 32 0 017 2 5 0 100 16 9 2 40 1 2 3 32 11 16 13428 PCT 1 2 x 1 8 0 017 2 5 0 100 16 9 2 40 1 2 1 8 3 4 13430 PCT 1 2 x 3 16 0 0...

Страница 24: ...1 2 x 3 16 PCT 3 4 x 1 8 PCT 3 4 x 3 16 PCT 1 x 1 8 PCT 1 x 3 16 PCT 1 1 4 x 3 16 Size H x W x D 85 x 80 x 10 6 mm 100 x 80 x 10 6 mm Weight 28 g 31 g Material polyamide polyamide Mounting four M5 scr...

Страница 25: ...nue of approximately 7 billion We deliver innovative equipment and components consumable supplies aftermarket parts software and digital solutions and support services through five operating segments...

Отзывы: