Quick guide for TBA firmware V1.070
7. Saving the recorded data to a USB memory stick.
Managing the recorded data is possible in the Logging and USB
Memory menu.
To select the Logging and USB Memory menu, press the [F1] – „Menu”
button on the home screen
(figure 31)
.
Settings and additional options menu.
(figure 32)
Using the [F2] button select the Logging and USB Memory menu and
then press the [F3] – „OK” button.
Logging and USB Memory screen
(figure 33)
allows you to manage
recorder banks. Select a bank by means of the adjuster.
Pink colour means that the Bank is full – there are recorded logging
data.
Green colour means that the data from the Bank have just been saved
to a USB memory stick.
Grey colour means that the Bank is empty.
There are 16 banks in total, each bank can contain 512 records.
Records are saved in the following time interval (the period of record
saving):
The frequency of record saving is set before the discharge process.
Transfer of data from the tester memory to a PC should be performed
by using a USB memory stick.
For USB memory sticks with the capacity up to 2 GB select the FAT16
file system while formatting them.
For USB memory sticks with the capacity over 2 GB select the FAT32
file system while formatting them.
The device does not support media formatted in the NTFS file system.
After placing the USB memory stick in the USB slot and locating the
cursor by means of the adjuster on a selected Bank, press the [F3] –
„Save” button. The data will be saved in a moment and the cursor will
change its colour to green.
Logging data files are saved in the CSV format – this format is supported
by popular spreadsheets.
As CSV is a text file in which the data are separated by a ‚;’ sign, you
can view its contents by using a text editor.
figure 32
figure 33
figure 31