Maintenance
55
10.1.7.1
Quick cleaning of the dust box
The dust box should be quickly cleaned daily or every 100 analyses at the latest:
Fig. 14: Quick cleaning of the dust box
Position
Component
1
Supplied brush
2
Top end of the filter housing
3
Bottom of the filter housing
4
Bottom O-ring at support
Grease is only applied to the bottom end of the filter housing (
3
) and the bottom O-ring (
4
). The
top end of the filter housing (
2
) and the O-ring on the top sealing mechanism should be clean
and remain absolutely free of grease.
Proceed as follows to quickly clean the dust box:
1. Ensure that the filter housing with the dust filter has been removed (see
”).
2. Using a suitable paper towel, remove the grease at the top end of the filter cartridge.
_. Clean the filter from the top end of the filter housing (
2
).
3. To do this, turn the supplied brush (
1
) in only one direction.
4. Collect the dust
in a suitable waste container
.
5. Insert the dust box again.
6. Turn the lock (see Pos. (
3
) in
”) 90° to the right.
7. Secure the connector on the metal filter heater.
The dust box has been cleaned.
1
2
3
4