3.29 Dust or soot from the resistance furnace
Pos : 3. 59 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 05Faults _C S-2000/ 0660 M odul D ust or s oot from the r esist ance f urnace @ 9\ mod_1425541504144_9. doc x @ 74477 @ @ 1
Fig 49: Front CS-2000
The sample has been inserted into the furnace too quickly:
–
See Operation Manual chapter
“
Applications / Coal analysis
”.
The sample weight is too high:
–
If a sample burns explosively, then there could be a shortage of oxygen. This can cause
some soot or oil vapors that pass through the hot zone of the furnace without burning. It
is necessary to reduce the weight of the sample, e.g. to 100mg or even to 50mg. If the
weight reduction doesn’t help, then operate differently by shifting the boat slowly towards
the beginning of the hot zone of the furnace. Leave the boat there and allow the sample
to start burning. Towards the end of the combustion, move the boat to the middle of the
hot zone, in order to obtain a complete combustion. See Operation Manual
“
Applications
/ Oil analysis
”.
The sample is metallic:
– Perform analysis by using the induction furnace. See Operation Manual.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...