5.7 Gas pump replacement
Pos : 7. 15 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-2000_s er vice/ 15 Ser vic e_C S-2000/1540 Modul G as pump r epl ac ement @ 9\ mod_1425543580653_9. doc x @ 74790 @ @ 1
Fig. 95: Gas pump
–
Switch off the analyzer and disconnect the mains power plug.
–
Disconnect plug (1) of the pump from the socket X4 on the board (2).
–
Disconnect the tubes (3) and (4) from the pump (5).
–
Remove the pump from the supporting panel (6) by pulling the four rubber buffers (7).
(If necessary they can be cut).
–
Reinstall a new pump in reverse order.
REMARK
For installing a new pump, the rubber buffers (7) of the new pump have to be pulled down from
underneath the supporting panel (6) by using slim pliers. If suitable pliers not available, the
board (2) can be removed by unscrewing the four nuts (8).
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...