Analysis
32
1.
The TIC module is placed between the furnace and the analyser:
2.
The outlet (15) of the furnace is connected to the connection (4) of the glass distributor
(3).
3.
The outlet (2) of the glass distributor (3) is connected to the upper fitting of the moisture
trap (13).
4.
The lower fitting of the moisture trap (13) is connected to the analyser inlet (14).
5.
The bottle of acid with dispenser (9) is placed to the right of the platform (8) and it is
connected to the connection (5) of the glass distributor.
6.
The glass distributor (3) with the glass bottle (6) are adjusted so that the whole surface of
the bottom of the glass bottle (6) is in touch with the surface of the heater (7).
7.
The heater (7) is switched on and the temperature is set at about 75°C (between setting 3
and 4).
8.
The stirrer is set to 400.
5.5.3 TIC-module Operation procedure
1.
Place the empty glass flask (
6
) on the balance.
2.
Press tare.
3.
Put the sample into the flask and enter the weight into the analyzer (F4-
button).
When powder sample sticks in the flask neck, add 2 ml of water (but don't transfer the
weight of the water!)
4.
Place a magnetic stirrer into the flask and attach the flask to the distributor (
3
).
5.
Raise the adjustable platform to support the flask. Check, and if necessary, readjust the
flask so that its bottom lies flat on the heated platform.
6.
Switch to TIC-mode with the TIC/TC toggle (16).
7.
Start analysis (F5 or click START).
8.
Inject acid in two or three doses
, when the word “Analyzing” appears.
9.
When all the CO2 has been released from the sample, the analyzer's signal will return to
the baseline level and the analysis will be terminated.
10.
Switch to TC-mode with the TIC/TC toggle (16).
NOTICE
The rotary speed of the stirrer should be kept low. The rotary speed and the acid dosing should
be done in a way to avoid sample particles being pushed up and stick on the inner glass
surface.
The heater must be switched on. Do not allow boiling or evaporation of the solution in the flask!
The table below shows approximate sample weight and acid volume depending on the
expected TIC content in the sample.
TIC-content
Sample weight
Acid
>5 %
100 - 200 mg
2×2 ml
1-5 %
200 - 500 mg
3×2 ml
<1 %
1000 - 2000 mg
3×3 ml
Used Acids:
Acetic acid 25% concentration or Phosphoric acid 50% concentration.
NOTICE
Only the carbon of easily decomposable carbonates can be determined. Carbonates which are
difficult to decompose cannot be measured. For example, elementary carbon (graphite, soot)
and cyanides cannot be analyzed.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...5 ...
Страница 63: ...Maintenance 63 Fig 35 Induction furnace Cone ...
Страница 80: ...Function description 80 Fig 51 Gas flow system ...
Страница 91: ...Miscellaneous 91 8 1 6 Furnace Fig 60 Furnace 71031 71010 ...
Страница 98: ...Miscellaneous 98 8 1 12 Tic module Fig 66 TIC module 35543 9008 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Страница 110: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...