74 - EN

liberoCONFIG

LI6002Eg

CALIBRATION

 6

6.3

Methods for Temperature Calibration

• 0°C Ice-water

Calibration, which uses the triple point of ice-water (0°C) as reference temperature. You 
can expect an accuracy of approx. 0°C ±0.1K.

• Calibration bath

If a calibration bath is used (-35°C .. 70°C), the reference sensor should be fastened to 
the data logger under calibration. This ensures, there is no temperature difference 
between the data logger and reference sensor. The watertight packed data logger 
should be immersed completely and the bath should be stable. Also make sure that the 
data loggers do reach the reference temperature. Repeated measuring and averaging 
can improve the measured values.

6.3.1

Ice Water Calibration Procedure

1. Fill up an insulated container, e.g. a camping coolbox with ice cubes. Use ice from an 

ice machine (-1°C) and not from the freezer (-20°C). Fill the coolbox with cold water up 
to the filling height of the ice. In order to mix it well, stir the contents.

2. The "Logging Interval" is factory set to 1 minute, the "Start Delay" to 50 minutes and the 

"Measuring Time" to 10 minutes.

3. Wrap up the data logger in watertight packaging material, e.g. latex glove.

4. Submerge the data logger completely in the ice water.

5. Carry out the calibration.

6. Evaluate the calibration.

 6.1 

View/Print Calibration

The calibration process lasts as long as the total time required for "Start Delay" + 
"Measuring Time" for each calibration point.

Содержание LIBERO Tx

Страница 1: ...LIBERO Tx PDF Logger Operation Manual...

Страница 2: ...mal conditions All main items of the software product comply with the operating instructions and the information in the help file The guarantee does not apply to consumables disposable batteries or an...

Страница 3: ...ystem requirements 14 3 3 Common liberoCONFIG 15 3 3 1 Start of liberoCONFIG 15 3 3 2 Options General 17 3 3 3 Options Additional 18 3 3 4 Menus 19 3 3 5 Buttons 19 3 3 6 Confirmation 19 3 3 7 Configu...

Страница 4: ...4 2 Error Messages 59 5 Example Transportation 60 5 1 Battery Lifetime Display 60 5 2 Transport Monitoring 61 5 3 Data Evaluation 61 5 3 1 Alarm Representation 61 5 3 2 Statistics 62 5 3 3 PDF report...

Страница 5: ...EN 5 liberoCONFIG LI6002Eg 7 2 ELPRO Customer Service Information 75 7 2 1 Software license 76 7 3 Revision History 763...

Страница 6: ...ture and humidity PDF report Logged transport information is created as pdf file liberoCONFIG Program for the configuration of a LIBERO LIBERO SmartStart Program for simple configuration of LIBERO Tx...

Страница 7: ...s can be configured directly with the LIBERO SmartStart application Trans fer is absolutely reliable and safe as no device settings are accessible during this process When assigning a profile with the...

Страница 8: ...CONF Mode on the respective USB interface Follow the instructions of the Windows hardware installation assistant Period of use 100 Days 400 Days 3 Years Sensor position Types 16 000 Temperature values...

Страница 9: ...your area to dispose of the data logger In any case it is recommended to contact the local EPA U S Environmental Protection Agency office http www epa gov International If possible dispose of the logg...

Страница 10: ...r portable electronic devices for use on airplanes in accordance with the definition in the Advisory Circular 91 21 1B of the Federal Aviation Administration FAA and were tested in accordance with RTC...

Страница 11: ...onfiguration Software 2 2 Config Mode when running Display shows data and status information To configure a PDF logger liberoCONFIG has to be used This software may be down loaded free of charge from...

Страница 12: ...er is put into the configuration mode it has to be connected to an USB port within 10 seconds Continue with 3 Configuration Software Configuration Mode Exit Press one of the buttons prior to connectin...

Страница 13: ...ration profiles to a large number of LIBEROs 3 1 Configuration file till SmartStart Pack Go exe file The configuration of an individual LIBERO Tx is carried out with liberoCONFIG This work is simplifi...

Страница 14: ...I6002Eg CONFIGURATION SOFTWARE 3 3 2 System requirements Windows XP 7 or 8 CPU 1 5GHz Memory 512 MB RAM Free hard disk space 100 MB Monitor 800 x 600 Pixel Information on the latest functions is given...

Страница 15: ...Start of liberoCONFIG On this screen all available LIBERO Txs are visible Start window of liberoCONFIG with 2 LIBERO Tx The following infor mation is shown LIBERO ID No Type Status Remaining battery...

Страница 16: ...tions Password length Specify the paper format for the reports 3 3 2 Options General Shows the information about the current program version and license number Refresh Refreshes the start window of li...

Страница 17: ...s can be made Options General Settings Language German English Spanish French Italian Dutch Default File Loca tions For the following profiles the file location can be defined or selected Configuratio...

Страница 18: ...the minimum password length View Print Template for Calibra tion Data Use built in template The factory defined layout of the calibration document is used for printouts Use custom template The custom...

Страница 19: ...up data loggers Save Profile Used to save a new profile in a file Apply The selected data loggers will be configured and started according to the open profile 3 3 6 Confirmation All successful actions...

Страница 20: ...by liberoCONFIG such as Description of the PDF report Logging functions PDF Options Format settings Handling options Configuration profiles are saved as LiberoCFG files can either be applied to multi...

Страница 21: ...figuration file Configuration or creation of a configuration file liberoCFG is only possible if at least one LIBERO Tx is shown in the start window Menu Device Edit Configuration The chapters 3 4 1 De...

Страница 22: ...aluation report up to 60 characters Additional Informa tion Free text that can be added to the profile to appear on the PDF report up to 8 lines of 80 characters PDF Filename Specified file name of th...

Страница 23: ...pending on the model the recording interval is between 1 or 5 minutes and up to 60 min utes www elpro com Logging Mode Logging is blinking Loop Logging is done continuously If the memory is full 16 00...

Страница 24: ...seconds Initial state Transit Arrived Initial State Depending on the transportation cycles you may switch several times between the two modes Start in Transit mode The LIBERO Tx starts recording with...

Страница 25: ...wn menu Selection Alarm disabled Alarm mode AN ALARM CAN BE RESET BY A NEW CONFIGURING OFF Alarm disabled No alarm monitoring takes place 3 4 4 Single Alarm Thresholds 3 4 5 Multiple Alarm Zones This...

Страница 26: ...ONFIGURATION SOFTWARE 3 3 4 4 Single Alarm Thresholds LIBERO Tx for temperature recording LIBERO THi1 x for temperature and humidity recording Upper Thresholds Lower Thresholds Data entry fields for t...

Страница 27: ...re fulfilled Used These check boxes are used to select the desired alarm zones Temperature T C or T F Data entry fields for the threshold values Alarm after An alarm is not triggered until the thresho...

Страница 28: ...for zone 3 and 5 accumulation deviations Graphical Representation Evaluation Temperature plot with zones Data Zone Tempera ture range C Add up time h Number of thresh old violations Plot section used...

Страница 29: ...rived period will be used for the statistical calculation If the ARRIVED button was not pressed all values until the moment of evaluation will be used Graphics all values in Memory Alarm handling Last...

Страница 30: ...Unique PDF report number comprising the device ID and the time stamp of the PDF report 2 File name specified for the PDF report during configuration 3 Date of report creation 4 Time of report creation...

Страница 31: ...I6002Eg CONFIGURATION SOFTWARE 3 3 4 6 3 Example PDF report possible hidden information Profile ID Not shown Alarm Conditions Not shown Alarm Conditions Alarm Status Not shown Alarm Condition Alarm St...

Страница 32: ...on UTC Daylight Saving Time 1 hour A set checkmark adds 1h to the selected time zone Default setting UTC Date Time Format Choose format settings for date and time Default setting DD MMM YYYY 31 Oct 2...

Страница 33: ...the TRANSIT button for 2 seconds Enables the buttons to mark the start and the arrival of the shipment Those markings are visible in the PDF report 3 4 6 2 Example PDF report Transit Arrived period Me...

Страница 34: ...d Display Statistics ONLY Measurement statistics shown on the display 5 3 2 Statistics Disabled Start Config only To start recording press the TRANSIT button for 2 seconds Status The buttons are inact...

Страница 35: ...g starts Use Start Delay Loop mode The delay time starts immediately after the LIBERO Tx has been configured Start Stop mode The delay time starts after the LIBERO Tx has been started by keypress Use...

Страница 36: ...play capability The Profil ID is always listed in the Device Configuration of the LIBERO PDF report 4 selection windows for defining the Profil ID By checking the box the Profil ID appears on the disp...

Страница 37: ...ation Profile All selected LIBERO Tx are directly configured with a previously saved profile 3 4 11 View Print configuration Creates a report with all configuration data This report contains as many p...

Страница 38: ...38 EN liberoCONFIG LI6002Eg CONFIGURATION SOFTWARE 3...

Страница 39: ...rd is set data transfer into elproVIEWER or elproLOG ANALYZE needs user authorisation Set Change Reset If New Password Repeat new Pass word are left blank the password will be reset 1 To reset the pas...

Страница 40: ...is used to validate a PDF report If the files have passed test results will be shown and can be printed or archived Procedure File integrity check 1 Select and open PDF reports that should be checked...

Страница 41: ...EN 41 liberoCONFIG LI6002Eg CONFIGURATION SOFTWARE 3...

Страница 42: ...ssibility to read the logged data from an already switched off LIBERO Tx as PDF report After the file has been read the LIBERO Tx switches itself off again automatically 3 6 3 Set Time Zone of Device...

Страница 43: ...carrier tracking number and viewable on the PDF report During the final programming process information can be added either in the form of plain text or dropdown menu All other critical device setting...

Страница 44: ...EN liberoCONFIG LI6002Eg CONFIGURATION SOFTWARE 3 Start LIBERO SmartStart Settings 3 7 1 Items 3 7 2 Configuration Report 3 7 7 Create SmartStart Pack Go File Return to liberoCONFIG File format libero...

Страница 45: ...EN 45 liberoCONFIG LI6002Eg CONFIGURATION SOFTWARE 3 3 7 1 Items Manage items Overview LIBERO SmartStart Settings...

Страница 46: ...e The title is the designation for the selected item Enter a short clear name Color Each item can be assigned a color The color serves to identify the item quickly at a later point while working with...

Страница 47: ...ent no access There are no entries possible If all entries are set to no access no entry prompts appear during profile assignment with SmartStart Pack Go In all other cases an entry window is opened t...

Страница 48: ...ents all configurations done by LIBERO SmartStart With the aid of the Placeholder Editor the used file name path and the parameters docu mented may be customized Reporting ON OFF Configuration report...

Страница 49: ...efault setting for saving the configuration report Subdirectories Filename This button is used to insert placeholders for the designation of subdirectories and file names 3 7 4 Report Content Report C...

Страница 50: ...ected placeholders determine the con tent of the PDF report 3 7 5 Placeholder Editor Example Placeholders are expressions that appear as variable information in the report Data logger ID This is the s...

Страница 51: ...ding to regional settings of the computer Time zone corre sponds to settings in the PDF report Configuration Time Format Configuration Time Format Time of the configuration of the LIBERO Tx with user...

Страница 52: ...ApplicationName Corresponds to the Window Title 3 7 7 Create SmartStart Pack Go Place holder Function dd Day with leading zero MM Month with leading zero MMM Shortcut month Jan Feb Mar MMMM Name of m...

Страница 53: ...Report opened with MS Excel Report opened with an editor Example of a report Line Description 1 Date and time configured with the placeholder Configuration time Format con figured 2 Text 3 Always use...

Страница 54: ...and configuration profiles File format exe Pack Go runs on any PC from a WEB FTP or file server or from the Internet 3 7 7 1 SmartStart Pack Go Settings Settings Program Window Window Title Window ti...

Страница 55: ...d LIBERO SmartStart settings 3 4 Configuring the LIBERO Tx 3 5 Passwords 2 Create an exe file which includes the following information Defined profiles and LIBERO SmartStart settings File name and tit...

Страница 56: ...tStart settings 3 7 LIBERO SmartStart Settings Mode The LIBERO Tx must be in OFF or ConF mode If the modes are not set properly a PDF report will be shown In the window the first item from the list is...

Страница 57: ...win dow for entering the free configuration data is opened The data can be entered with the keyboard or with the Barcode reader Key to entries 1 Edit text 2 Add dropdown 3 Insert dropdown 4 Add text...

Страница 58: ...the LIBERO Now disconnect the LIBERO from the USB port and continue with the next LIBERO 4 1 Change of Settings A SmartStart Pack Go can contain several items profiles Click the setting window with t...

Страница 59: ...MARTSTART PACK GO 4 4 2 Error Messages In case of an error a message will be displayed Reasons for error massages Profile incompatible Cancel has been during data configuration problems LIBERO has bee...

Страница 60: ...m Bat End is blinking Remaining lifetime 30 days If no lifetime for Bat End is displayed change LIBERO Tx at your next possibility as soon as possible Battery operation time has been expired After 20...

Страница 61: ...product related transportation SOP Data logging LIBERO Tx have several possibilities to start stop a recording 3 4 2 Logging 5 3 Data Evaluation 5 3 1 Alarm Representation To enable this function 3 4...

Страница 62: ...shown For an immediate exit press the TRANSIT button again or wait approximately 10 seconds till the LIBERO Tx displays measurement values again P Id Depending on the pro file settings the Profil ID...

Страница 63: ...t of the LIBERO Tx Exit statistic display If the last statistic display has been reached return to standard display by pressing the ARRIVED button or wait approximately 10 seconds till the device disp...

Страница 64: ...r ARRIVED button until the end of the menu or wait approximately 10 seconds till the device displays measurement values again P Id Abbrevation for Profil ID Alarm zones ZZ 1 to 5 Time within temperatu...

Страница 65: ...rinted above the graphic in bold WARNING Sensor error If a sensor short circuit or disconnection was detected this warning is printed above the graphic in bold LIBERO Tx have the feature to create a P...

Страница 66: ...off device 3 6 2 Create PDF from Switched Off Device For data archiving or later evaluation the PDF file may be copied to any desired drive elproVIEWER and elproLOG ANALYZE For further evaluation LIB...

Страница 67: ...3 3 4 Menus 2 Set LIBERO Tx into the configuration mode 2 3 Start Configuration Mode 3 Carry out the calibration 6 1 View Print Calibration Shows a calibration report This report contains as many page...

Страница 68: ...een calibrated according to the requirements of a single or multi point calibration procedure Document Calibration Certificate After the calibration the docu ment shows true and rated values instead o...

Страница 69: ...2Eg CALIBRATION 6 6 2 New Calibration Starts the process for calibrating a LIBERO Tx 6 2 1 Calibration points Used to determine the number of calibration points and the calibration values LIBERO Tx fo...

Страница 70: ...4 calibration points selectable Calibration values Data entry fields for each calibration value Rated Value and the Allowed Tolerance Calibration data could be saved as a calibration profile To preven...

Страница 71: ...humidity recording Process Start up delay Data entry field for the desired delay time between manual calibration start and start of data logging The Start Delay depends on the time required to reach...

Страница 72: ...t Calibration 6 2 3 Start of Calibration To start to calibration press Apply Waiting for start After the data logger has been configured it is waiting for the start of the calibration CAL x Calibratio...

Страница 73: ...ding to the Allowed Tolerance For the next calibration point press the TRANSIT or ARRIVED button 6 2 1 Calibration points End of calibration The display changes to ConF Mode To print the calibration d...

Страница 74: ...mperature Repeated measuring and averaging can improve the measured values 6 3 1 Ice Water Calibration Procedure 1 Fill up an insulated container e g a camping coolbox with ice cubes Use ice from an i...

Страница 75: ...mation If you need any assistance from the ELPRO Customer Center please make sure you can provide the following information Software version Select About in the menu option Help Device type used Which...

Страница 76: ...icense information User name xxxxx xxxxx Company name xxxxx Serial number 8197 7949 0717 License agreement This license entitles the owner to install the software on several computers if those belong...

Страница 77: ...ase 8 9 CE 10 Checksum 20 36 Color 46 Configuration Mode 12 Configuration Mode Exit 12 Configuration password 39 Configuration report 37 Copy 46 Cumulative events 27 D Data access password 39 Data eva...

Страница 78: ...Reset code 39 Reset password 39 S s e 65 Save profile 19 Selection 46 Sensor error 65 Single alarm 27 Single events 27 SmartStart 43 75 SmartStart configuration 46 SmartStart configuration report 48 S...

Страница 79: ...EN 79 liberoCONFIG LI6002Eg MISCELLANEOUS 7...

Страница 80: ...Operation Manual liberoCONFIG LI6002Eg 09 2013 ELPRO BUCHS AG Lang ulistrasse 62 CH 9470 Buchs SG Switzerland E Mail swiss elpro com For local agencies see www elpro com...

Отзывы: