Contact inputs D1 and D2
7. Contact inputs D1 and D2
Assign
Defrost input
The status of D1 and D2 is not recorded in the alarm protocol and has no influence on the alarm contacts.
They are first logged to memory on the datalogger at the end of the following recording interval in memory.
In the measurement table elproLOG ANALYZE both contacts are represented as D1, D2 or MarkPos.
Alarm forwarding
Alarm forwarding is used in combination with the elproMONITOR software. The status (alarm / no alarm)
of the contacts is checked during the update of the monitor data. Status changes between two updates
are not detected.
Both inputs are busy with dual functions!
In each case, only one function should be used!
Function
possible configurations
Assign
D2 key
D2 key
Defrost input
D1 external
D1 external
Alarm forwarding
D1 external
D2 external
D1 external and D2
external
D2 key
See function F2, timestamp designated as D2 F
This function can be used to register incidents such as a watchman's patrol on the logger.
D1 external
As defroster input (
Threshold value and alarm function
) or for alarm forwarding.
Wiring
Supply, Contact input, Alarm 2
D2 external
For alarm forwarding.
Wiring