Contact inputs D1 and D2

7. Contact inputs D1 and D2

Assign
Defrost input

The status of D1 and D2 is not recorded in the alarm protocol and has no influence on the alarm contacts. 
They are first logged to memory on the datalogger at the end of the following recording interval in memory. 
In the measurement table elproLOG ANALYZE both contacts are represented as D1, D2 or MarkPos.

Alarm forwarding

Alarm forwarding is used in combination with the elproMONITOR software. The status (alarm / no alarm) 
of the contacts is checked during the update of the monitor data. Status changes between two updates 
are not detected.

Both inputs are busy with dual functions!
In each case, only one function should be used!

Function

possible configurations

Assign

D2 key

D2 key

Defrost input

D1 external

D1 external

Alarm forwarding

D1 external

D2 external

D1 external and D2 
external

D2 key

See function F2, timestamp designated as D2 F

4. 

Menu

This function can be used to register incidents such as a watchman's patrol on the logger.

D1 external

As defroster input (

6. 

Threshold value and alarm function

) or for alarm forwarding.

Wiring 

8.3 

Supply, Contact input, Alarm 2

D2 external

For  alarm forwarding.
Wiring 

8.3 

Supply, Contact input, Alarm 2

Содержание ECOLOG-NET LP4

Страница 1: ...ECOLOG NET LP4 EL6006Ee ECOLOG NET LP4 Operation Manual...

Страница 2: ...r normal conditions All main items of the software product comply with the operating instructions and the information in the help file The guarantee does not apply to consumables disposable batteries...

Страница 3: ...ng mode 8 2 8 Overlaying graphical data 8 3 Configuring the datalogger 9 3 1 Assign network address 9 3 1 1 Desktop installation 9 3 1 2 Documentation 9 3 2 Installing the datalogger 9 3 2 1 Installat...

Страница 4: ...10 Status and error messages 24 10 1 Datalogger display and elproLOG ANALYZE 24 10 2 Datalogger status in elproLOG ANALYZE 24 11 Maintenance 25 12 Technical data 26 12 1 Measuring ranges and Accuracy...

Страница 5: ...ct to CE marking The manufacturer guarantees that this product complies with the relevant guidelines see 13 Declarations of conformity This product must be disposed of in accordance with WEEE Waste el...

Страница 6: ...ft side of the housing and on the front a display and a membrane keyboard 1 Network connection 2 Connection for 4 Pt100 sensor 200 C 600 C 3 USB connection 4 Power supply contact inputs and alarm outp...

Страница 7: ...anteed that the loggers will function properly if the datalogger is exposed to tempera tures which exceed the specified threshold range Experience has shown that the battery freezes at approx 50 C tha...

Страница 8: ...y The datenlogger ECOLOG NET has a power save mode which switches off the display As a result mea surements are only made during the defined log interval Four small circles located in the display indi...

Страница 9: ...igned to the datalogger during this workstep This work should be per formed at the place of use prior to final installation e g in an office 3 1 2 Documentation Documentation for the implemented confi...

Страница 10: ...g the datalogger 3 2 5 Verification of the installation Check the installation and ensure that sensor positions alarm parameters and network addresses are cor rect Test connectors can be used as an ai...

Страница 11: ...rm or when the self sustaining function is selected and an alarm is registered 6 2 Window Setup of Alarm parameters Jump to menu item 5 Reset alarm display and alarm contacts as well as entry of the r...

Страница 12: ...ly Definition of the alarm thresholds 6 Threshold value and alarm function Reset alarm This function is used to acknowledge an alarm mes sage Set date and time Used to adjust the integrated datalogger...

Страница 13: ...Datalogger setup 6 1 Conditions Threshold values Alarm Delay Time Alarm output The measured value must be outside the defined tolerance range i e the measured value is higher than the maximum allowab...

Страница 14: ...nction Alarm thresholds Input fields for the lower and upper threshold values Hysteresis The Hysteresis is used to avoid fluttering of the alarm contact A change of state in the contact only occurs wh...

Страница 15: ...semiconductor switch break contact to datalogger GND Semiconductor contact alarm 2 Only reacts when there is an alarm This is a semiconductor switch make contact to datalogger GND Internal Buzzer Ser...

Страница 16: ...1 cycle time 2 Alarm is logged ECOLOG NET Alarm contact responds immediately ANALYZE 2 Alarm measurement ANALYZE Alarm protocol with 2 entries alarm on off MONITOR Represents an alarm for 7 cycle time...

Страница 17: ...alarm and therefore it is not recognized as an alarm 5 Reset of delay time by short threshold value violations ECOLOG NET Alarm contact does not react Even in the case of repeated short threshold valu...

Страница 18: ...of the contacts is checked during the update of the monitor data Status changes between two updates are not detected Both inputs are busy with dual functions In each case only one function should be...

Страница 19: ...ction 8 1 Pt100 sensors 1 4 Pt100 Connector metal lized DB9 female Part no 800504 Schema Pt100 3 wires Pt100 4 wires cable length 12m Part no ECA_PT100_SILxx Pt100 4 wires cable length 12 30m Part no...

Страница 20: ...20 EN ECOLOG NET LP4 EL6006Ee Pin assignments and connections Pt100 4 wires cable length 30m Part no 800520 CABLE LENGTH 12 1 Measuring ranges and Accuracy...

Страница 21: ...logger operating volt age 6 sc Alarm contact relay contact 7 sa Alarm contact relay contact 8 Gnd Ground 9 Gnd Ground Alarm contact relay contact Relay contact represented in dead state Connection sa...

Страница 22: ...und 5 V in External power supply logger operating volt age 6 Gnd Ground 7 Gnd Ground 8 Gnd Ground 9 Gnd Ground FOLLOW THE SAFETY AND APPLICATION INSTRUCTIONS OF THE POWER SUPPLY UNIT Contact input Ala...

Страница 23: ...or VDC 500mA Alarm Threshold value violation or cable break Alarm Contact open Datalogger does not require external power supply Semi conductor contact Use only for DC voltage DCV Switch load max 24VD...

Страница 24: ...of the batteries is displayed in the elproLOG ANALYZE status information Battery replacement 11 Maintenance Firmware The version of the firmware is documented in the status Display elproLOG ANALYZE CA...

Страница 25: ...ium 3 6V 2100mAh AM3 LR6 AA Battery An energy consumption counter is used to monitor datalogger battery life For this reason only the specific manufacturer recommended battery should be used Do not re...

Страница 26: ...1 C 0 2 C 400 1 C 500 0 C 0 1 C 0 3 C 500 1 C 600 0 C 0 1 C 0 5 C This information applies to the ECOLOG NET LP4 sensors at room temperature and without Pt100 Length Cable typical errors 50m 4x0 25mm2...

Страница 27: ...EN 27 ECOLOG NET LP4 EL6006Ee Technical data IT ES EN EN FR IT ES 12 3 Dimensional view 1 Battery for a battery exchange the rear wall has to be removed 2 Screws for the attachment of the rear wall...

Страница 28: ...ension cable for Pt100 sensor 0 5m for cable guides e g refrigerators ECA_4C_xxM_M12 Connection cable Pt100 up to 30m with M12 plug ECA_LIYCY4X0 34 4 wire connection cable with M12 plug for cable leng...

Страница 29: ...EN 29 ECOLOG NET LP4 EL6006Ee Declarations of conformity IT ES EN EN FR IT ES 13 Declarations of conformity...

Страница 30: ...er 04 03 2011 First edition replaces EL6003Bk A Gubler 21 12 2011 a Minor changes A Gubler 01 10 2012 b New Graphic A Gubler 03 06 2013 c No marking of the sensor cable required A Gubler 13 01 2014 d...

Страница 31: ...ation 9 E ECOLOG NET LP4 6 elproLOG ANALYZE 9 24 elproLOG CONFIG 9 elproLOG MONITOR 9 Environmental conditions 7 Error messages 11 24 Explosion protection 6 Extended setup 12 Extended setup 12 F Firmw...

Страница 32: ...Operation Manual ECOLOG NET LP4 EL6006Ee 11 2016 Head Office ELPRO BUCHS AG Lang ulistrasse 45 9470 Buchs SWITZERLAND e mail swiss elpro com For local representatives see www elpro com...

Отзывы: