All the contents aspired reach the container via aspiration hose.
Reusable Collection Container
The reusable collection container is fixed on the side of the motor unit. The bacteria filter between the device and the collection
container prevents bacteria and liquid droplets from escaping into the motor unit and scattering the environment through the vibration
damping element. When the bacteria filter gets wet, any contamination or discoloration should be urgently replaced. If the device is used in a
patient at risk of contamination, the bacteria filter should also be replaced.
Never wash, sterilize or autoclave the bacteria filter.
Important!
An overfill prevention apparatus prevents the escape into the motor unit. A closing plug floating on the surface closes the outlet
when the secretion reaches a certain level.
Voltage Supply
Voltage Supply can be provided as follows:
With an internal battery pack,
By using a charging connection cable, over an existing 12 V vehicle electrical installation
With adapter/charger.
The capacity indicator shows the charge status of the battery pack. When you connect the device to an external voltage supply,
the battery pack starts charging automatically.
3. Safety Information and Warnings
For both your own and your patients safety and in accordance with the demands and requirements of the 93/42 / EEC directive, please
consider the following:
Please use LIFETIME SA-PB device only for its intended purpose.
Please read these instructions for use carefully. The instructions for use are part of this device and should always be available
Before working with the LIFETIME SA-PB device, you should know the necessary information about the use of the device.
Use the LIFETIME SA-PB device only if you have sufficient knowledge of medical education and suction technique. Improper
use and practices can lead to serious bodily harm
During suction, be careful not to cause any injury to the patient's mouth and sink area, especially mucous membranes.
During bronchial suction procedure, operate the device by obeying the sterility rules and use only sterile suction catheter.
When operating with the multi-use collection container, the device must always stand upright, because the ball of the overfill
assembly cannot safely close the connection to the suction pump and secretions may enter the suction pump. This can cause
damage to the pump.
The device must be in an upright position once the collection container is used. This prevents liquids or secretions from
spreading into the overflow valve filter, which is integrated into the once used suction bag. If you cannot prevent this, the
bleed feature of the overflow valve filter will be lost and you will need to replace the once used suction bag
We recommend that you have an alternate suction possibility in reserve, due to the possibility of device failure.
To prevent an infection or bacteriological contamination, Pay attention to the part 6 "Hygienic Preparation Procedures"
section.
If foreign materials are used, there may be dysfunctions and limited availability. In addition, the requirements for
biocompatibility may not be met. In such cases, be aware that if neither the accessories recommended in the leaflet nor the
original spare parts are used, any warranty and liability provisions will be completely void.
Please let the all audits and repairs be performed by manufacturer, ELMASLAR or its specialized staffs.
The battery pack should never be over-discharged in order to increase the lifetime of the battery pack. Please charge the
battery pack at the latest when the red 10% LED of the capacity indicator lights up.
To avoid the risk of over-discharging, never store LIFETIME SA PB in an uncharged state. First charge the battery pack
properly
Battery packs are also discharged when not in use. The rate of self-discharge increases in direct proportion to the temperature,
and this is about 50% when it is not used at 20 ° C for 1.5 months. Therefore, pay attention to the function control ranges.
Never insert the bacteria filter into the disinfection liquid, otherwise the properties of the filter will be lost.
Motor unit should not be sterilized in autoclave.
Do not immerse the motor unit in a disinfection solution.
Do not use wireless phone (mobile phone) nearby LIFETIME SA-PB. Even if a mobile phone is used in the driver's cabin,
using the LIFETIME SA-PB device in the patient compartment of the ambulance does not cause any problems.