Senko Multi Gas Detector
26
3.3. Alarm Display
Type
Set-Off Condition
LCD Display
Alarm Sound & Vibration Display
LOW Alarm
Exceed LOW alarm
value
icon & gas
concentration levels
displayed
When the user presses
Key after noticing that the LOW alarm sets off, the
sound stops, but the vibration and LED alarm remain.
HIGH
Alarm
Exceed HIGH alarm
value
icon & gas
concentration levels
displayed
The user must leave the area immediately, and the sound alarm/vibration/LED alarm
stops when the device goes to a safe area where the concentrations are normal
TWA Alarm
When exceeding TWA
alarm value
icon & gas
concentration levels
displayed
The alarm sets off when the hourly average levels of the gas concentration for the
last eight hours exceed the TWA concentration, and the sound alarm/vibration/LED
alarm stop when the gas concentration levels reach the alarm set-off value as the
user goes to a safe area.
STEL
Alarm
When exceeding STEL
alarm value
icon & gas
concentration levels
displayed
The alarm sets off when the hourly average levels of the gas concentration for the
last 15 minutes exceed the STEL concentration, and the sound alarm/vibration/LED
alarm stop when the gas concentration levels reach the alarm set-off value as the
user goes to a safe area
SENKO IR-LINK (Options)
Bump Test
Request Date for
Bump Test
Stops after Bump Test
Bump Test Interval: Notices the user on a regular basis to check the device.
Execute
Calibration
Request Date for
Calibration
Stops after Calibration
Calibration Interval: Notices the user on a regular basis to calibrate the sensor.