4 DIRECTIONS FOR USE
4.1 DC VOLTAGE MEASUREMENT
1. Switch the calibrator on using the On/Off button (
5
).
2. Press the input/output selection key (
8
) to display "INPUT" (
13
).
3. Press the V mV selection key (
6
) to select V DC or mV DC (
20
) according
to the desired measurement range.
4. Connect the red test lead to the V terminal (
4
) and the black lead to the
COM terminal (
3
).
5. Connect the other end of the red test lead to the positive pole of the
voltage to be measured and the black lead to the negative pole (earth).
6. The result is displayed on (
16
).
* The numbers in parenthesis refer to the Description of the front panel (Page
21) and the Description of the display screen (Page 22).
4.2 DC VOLTAGE OUTPUT
1. Switch the calibrator on using the On/Off button (
5
).
2. Press the input/output selection key (
8
) to display (
14
) "OUTPUT".
3. Press the V mV selection key (
6
) to select V or mV (
20
) according to the
desired output voltage.
4. Press the adjustment keys (
9
), (
10
), (
11
), (
12
), to set the desired value.
5. Connect the red test lead to the V terminal (
4
), and the black lead to the
COM terminal (
3
).
6. Connect the other end of the red test lead to the positive pole of the voltage
to be measured and the black lead to the negative pole (earth).
7. If you want to change the output value or range, press the setting
adjustment keys (
9
), (
10
), (
11
), (
12
) or the V mV selection key (
6
).
4.3 DC CURRENT MEASUREMENT
Measurement of an external current.
1. Switch the calibrator on using the On/Off button (
5
).
2. Press the input/output selection key (
8
) to display (
13
) "INPUT".
3. Press the mA
–mA% (
7
) selection key to select mA or mA% (
19
)
according to the desired display.
In the mA% mode, the 4-20 mA value will be displayed in the secondary
display area (
22
).
4. Connect the red test lead to the mA terminal (
2
), and the black lead to
the COM terminal (
3
).
5. Connect the other end of the red test lead to the positive pole of the
current to be measured and the black lead to the negative pole (earth).
6. The result is displayed in (
16
) or (
22
).