Elma 6100EVSE
23
Maintenance
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or
solvents. If the meter is stored away for 60 days or more, remove the battery and store it
separately.
Battery replacement
1. Remove the Phillips head screw that secures the rear battery cover (backside upside)
2. Replace Two "AAA" 1.5V Alkaline Battery.
3. Secure the battery cover.
Fuse replacement
The instrument is secured with 2 fuses for protection. When replacing, use the same type of fuses.
1. Remove the Phillips head screw that secures the fuse cover (backside down)
2. Replace the failed fuse. Type F10A/600V or Type F600mA/600V
3. Secure the fuse cover
General specification
Recommended calibration interval: 1 year
Display
6000 counts LCD
Continuity check
Threshold 50
; Test current < 0.5mA
Diode test
Test current typical 0.3mA; Open circuit typical voltage < 3.3VDC
Low battery indication
is displayed
Over-range indication
‘
OL
’ is displayed
Measurement rate
2 readings per second, nominal
Input Impedance
10M
(VDC and VAC)
AC response
True rms (50-400Hz, AAC - VAC and Fleksibel AC strømtang)
Operating Temperature
5
o
C to 40
o
C
Storage Temperature
-20
o
C to 60
o
C
Operating Humidity
Max 80% up to 31
o
C decreasing linearly to 50% at 40
o
C
Storage Humidity
<80%
Operating Altitude
Max. 2000 meter
Battery
2 Psc. " AAA" 1.5V Alkaline battery
Fuses
1 Type F10A/600V og 1. Type F600mA/600V
Auto power OFF
After approx. 15 min.
Dimensions & Weight
121 x 67 x 35mm; 190g
Safety
For indoor use and in accordance with the requirements for double
insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage
Category III 600V ,Pollution Degree 2
WARNING:
To avoid electrical shock, disconnect the meter from any circuit, remove the test
leads from the input terminals, and turn
OFF
the meter before opening the case for
change of battery or fuses. Do
not
operate the meter with an open case.