background image

ELK-319PB1 Heavy Duty 1 Button Panic

Description

The ELK-319PB1 is a Wireless Heavy Duty 1 Button Panic 

device that that transmits to an alarm panel when pressed.  It 

can be used police, medical, etc. alarm activation.  

The device also transmits low battery (as needed) and is 

powered by a single (1) replaceable 3VDC lithium coin-cell 

battery.

This sensor is compatible with Elk’s 319MHz Receivers/Panels 

as well as many other panels that operate on the 319.5MHz 

Frequency and adhere to the ITI/Interlogix protocol.  

Specifications

RF frequency:  319.5 MHz 

Compatibility:  ELK-319 Receivers/Panels & other panels  

 

that operate on the 319.5MHz Freq. and 

   

adhere to the ITI/Interlogix protocol 

Battery type:   One (1) 3VDC lithium coin-cell battery

   

(Panasonic or Varta Model CR2450)

Operating

Temp Range:    32 to 120°F (0 to 49°C) 

Storage

Temp Range:   -30 to 140°F (-34 to 60°C) 

Relative

Humidity:  

95% non-condensing 

Dimensions      3.0” x .2.0” x 0.7” in. (L x W x D)

Accessories:  Belt Clip, Lanyard Clip, Wall Mount Bracket 

Programming (Enrollment)

The following is a general guideline for programming (enrolling) 

a sensor into the receiver or panel. For more extensive 

instructions please refer to the receiver or panel instructions. 
1. Power up the device by 

pulling and removing the 

battery isolator tab.

2.  Place the panel into the 

Program mode. 

3.  Proceed to the WIRELESS 

SETUP menu. 

4.  Select the appropriate zone/sensor location number. 
5. Start the panel sensor Enroll process.  When the panel 

prompts to trip the sensor for learning do the following: 

  -  Press the Panic Button.  The panel should acknowledge 

the sensor has been learned by keypad display and/or audio 

alert (depending on the panel).  If enrollment fails to occur 

try repeating the process.  

6.  Repeat the above process for any additional wireless 

sensors.  Proceed to the zone programming to assign each 

sensor’s zone definition.  

7.  Exit programming mode when finished.

Testing the Sensor 

It is recommend that all wireless sensors be thoroughly tested 

after installation and programming.

Wireless Testing

Wireless testing provides a baseline for the distance the panic 

sensor will nominally operate away from the control or receiver.
1.  Place the panel into the Walk Test mode.
2.  Monitor the keypad display while pressing the button.  Refer 

to the receiver or panel for complete instructions.  In some 

cases the panel may provide a signal strength indication.

3.  Walk away from the panel or receiver and test at various 

locations making sure the panel acknowledges.

4.  Exit Walk Test mode when finished.

Operational Testing

If the alarm system is centrally monitored then always notify 

them prior to operational testing.

1.  Press  and  hold  the  panic  button  until  the  specific  alarm 

reaction (audible or silent dial report) begins.

2.  Disarm the alarm panel to silence or reset the alarm.

Accessory Attachment Options

There are three (3) options for carrying the Panic Button.  

TM

  Interlogix is a registered trademark belonging to United Technologies.

Belt Clip - allows Panic Button 

to be clipped onto a belt or 

waistband for handsfree carry 

and ease of access.

Wall Mount Bracket - used in 

conjunction with the belt clip, this 

wall or cabinet mounting bracket 

allows the Panic Button to be 

attached to a fixed object yet still 

be removed for portability.

Lanyard Clip - used in place of the belt clip to allow Panic Button 

to hang from a lanyard (not included).

Note: The belt clip must be installed on the panic device to 

fit into the bracket

Note: Tighten 

the bottom case 

screw after 

removing battery 

tab.

Battery isolation 

tab.

Отзывы: