ELIS PLZEŇ a. s.
Project design, installation and service manual
Page 18 o
f
33
Electromagnetic flowmeters of the type series
FLONEX FXx11x and FLONET FH30xx
Meter control instructions
ELIS PLZE
Ň a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722
Es 90 686 K
The function selected by buttons
► and ▲ will appear backlighted in yellow on the last display line.
Confirm the selected by
ENTER
.
Upon selection of item
Function negation
the output signal will change sign. The other functions
of the multi-function output will remain as they are.
5.1.4.4.1. Frequency for Q+
Using buttons
► and ▲, set the flow rate in litres per second corresponding to the output frequency of
1kHz. Confirm the set value by
ENTER
.
The display will return
to the previous menu item (
Flow rate at
1kHz
).
Return to Main menu:
3 x ESC
Return to Basic screen:
ESC 2 s
The flow rate value Q+ for 1kHz shall always be set in litres per second. Conversion of the volume
flow rate in litres per second to flow rates in other units is described above in the procedure for
setting the
100%
value.
5.1.4.4.2. Frequency for
Q−
See the procedure for setting the frequency for Q+.
5.1.4.4.3. Frequency for |Q|
See the procedure for setting the frequency for Q+.
Output 1 function
Frequency for Q+
Flow rate at 1kHz
Flow rate at 1kHz
l/s
+ 1 . 00
ENTER
ENTER
ENTER