5.1.4.2.4.
Backlighting time
Setting range: 0 to 999s
Value 0 is reserved for unlimited backlighting time.
The flashing underline marker shows the digital position ready to be set.
Set the display backlighting time using buttons
► and ▲ and confirm the setting by actuating the
ENTER
button. The display will return to the previous menu item (
Backlighting time
).
Return to Main menu:
1 x ESC
Return to Basic screen:
ESC 2 s
To actuate the “sleeping” display, depress any of the four control buttons.
5.1.4.2.5.
100%
By selecting the
100%
item, the user may define the flow rate to which the instantaneous flow rates will be
related (in per cent); the same applies to the bar graph display.
The flashing underline marker shows the digital position ready to be set.
Using buttons
► and ▲, set the flow-rate value in litres per second corresponding to the intended
maximum (100%) flow rate. Use button
► to move the position marker, and button ▲ to set the numeric
value 0 to 9. Confirm the setting by depressing the
ENTER
button. The display will then return to the
previous menu item (
100%
).
Return to Main menu:
1 x ESC
Return to Basic screen:
ESC 2 s
The maximum flow rate value shall always be set in litres per second irrespective of the
currently used flow rate unit. The flow rate value set as 100% shall not, in consideration of
the meter sensor size, correspond to the fluid flow velocity in excess of 10 meters per
second. It is recommended to set the maximum flow rate corresponding to the fluid flow
velocity between 3 and 5 m/s.
Volume flow rates in litres per second corresponding to various flow rate units
Unit
(litres per sec.)
l/s
1.00
l/min
1.666667E-02
l/h
2.777778E-04
Mil L/day
1.01
Backlighting time
s
000
100%
l/s
+1.01