CENTRA-R 60/120 -US 

Operator Manual

Содержание CENTRA-R 120 US

Страница 1: ...CENTRA R 60 120 US Operator Manual...

Страница 2: ......

Страница 3: ...reproduction and use apply to all media in which this information may be placed VWS UK Ltd pursue a policy of continual product improvement and reserve the right to alter without notice the specifica...

Страница 4: ...e Operational day screen 030 19 4 2 Continuous Recirculation 24 7 screen 029 19 5 MAINTENANCE 20 5 1 Replacing composite vent filters LC136 20 5 2 Installing an optional Labpure L1 purification cartri...

Страница 5: ...ELGA CENTRA R 60 120 US Operator Manual CENTRA R 60 120 US MANU40596 Version 02 01 16 Page 3 9 TECHNICAL SPECIFICATION 42 10 WARRANTY CONDITIONS OF SALE 44 11 USEFUL CONTACT DETAILS 45...

Страница 6: ...THIS APPLIANCE MUST BE EARTHED It is essential that the electrical supply to the CENTRA R 60 120 US is isolated before any items are changed or maintenance work performed The external isolator providi...

Страница 7: ...k L1 cartridge 6 months 2 years LC177 Protek L2 cartridge 12 months 2 years LC181 E cartridge 36 months 2 years LC187 Labpure L1 cartridge 3 to 12 months 2 years LC204 Micro filter cartridge 12 months...

Страница 8: ...lows the unit to be quickly installed and commissioned It is recommended that a trained ELGA representative carry out the installation of the product 2 2 Commissioning The unit is supplied with the so...

Страница 9: ...IGHT Mute Alarm Down CAUTION If the Master Key is lost a new Key can only be programmed by an ELGA Service Engineer 3 1 User Recognition Keys Your unit is supplied with the following User keys Master...

Страница 10: ...operation Note The User Key does not prevent access to the PROCESS function in case of emergency 1 SWITCH on the main electrical supply to start the controller set up sequence This takes several seco...

Страница 11: ...RESS Step 4 Audible alarm enabled disabled screen 020 This screen provides the option of either turning on the audible alarm causing it to sound while the alarm icon flashes or turning off the audible...

Страница 12: ...ater quality alarm will sound This alarm does not stop RO production but can extend or start flush routines 1 PRESS to select alarm point 0 20 30 40 50 or 100 S cm 2 PRESS to proceed to Step 8 3 If 0...

Страница 13: ...e alarm does not stop the unit and will automatically reset if the temperature returns below the set point 1 PRESS to select alarm point in increments of 1 ranging from 20 C to 50 C OR PRESS to procee...

Страница 14: ...eration is selected ON Step 13 Periods of operation screen 030 For efficiency and to reduce the likelihood of heat build up the normal operational hours can be selected During OFF periods the unit wil...

Страница 15: ...ng the selected off periods the unit can be restarted by pressing PROCESS A few minutes operation should be allowed before use to allow optimum water purity to be reached Step 15 Display viewing angle...

Страница 16: ...ue to the correct reading 2 PRESS to accept Step 18 Data output screen 035 The CENTRA R 60 120 US can transmit operational data to an Xport storage device 1 PRESS to change ON OR PRESS to proceed to S...

Страница 17: ...keys are registered maximum of 6 users 5 PRESS to complete settings Step 21 Programming of Sanitization Keys screen 038 Sanitization Keys are required to start a Sanitization These keys can be delete...

Страница 18: ...s screen is only applicable when a Labpure L1 Cartridge LC187 is installed 1 PRESS button to reset cartridge replacement date OR PRESS to accept replacement date and proceed to vent filter replacement...

Страница 19: ...SS button to reset Protek L1 L2 replacement date OR PRESS to accept replacement date and proceed to Ultra filter replacement 2 PRESS confirm that resetting is required OR PRESS to abort reset 3 PRESS...

Страница 20: ...0 US MANU40596 Version 02 10 16 Step 7 Sanitization reminder 1 PRESS button to reset Sanitization reminder date OR PRESS to accept reminder date and complete settings 2 PRESS confirm that resetting is...

Страница 21: ...water temperature During the sleep period the unit will display It is possible to override this mode by PRESSING 4 2 Continuous Recirculation 24 7 screen 029 Refer to Section 3 2 Step 12 If the unit...

Страница 22: ...filters LC136 The Composite Vent Filters CVFs should be replaced in the following circumstance When indicated by the alarm screen prompt or after a maximum of six months 1 ENSURE process is OFF and IS...

Страница 23: ...front doors and LOCATE bypass block LA698 as shown 3 PUSH bypass block FORWARD then LIFT and finally PULL to remove Store safely within unit for future use 4 REMOVE Labpure L1 from packaging 5 REMOVE...

Страница 24: ...doors and LOCATE Labpure L1 cartridge 3 PUSH Labpure L1 FORWARD 1 then LIFT 2 finally PULL 3 to remove and discard the cartridge 4 REMOVE new Labpure L1 from packaging 5 REMOVE sealing plugs from inl...

Страница 25: ...release retaining arm 3 4 SWING retaining arm 3 fully open 5 SLIDE by pass block 4 forward to remove Store safely within unit for future use 6 REMOVE Micro Filter from packaging 7 REMOVE sealing plug...

Страница 26: ...CATE filter as shown 3 LIFT wing 1 and rotate anti clockwise 4 PIVOT latch 2 to release retaining arm 3 5 SWING retaining arm 3 fully open 6 SLIDE filter 4 forward and discard 7 REMOVE new filter from...

Страница 27: ...RCURY MAY CAUSE REDNESS OR IRRITATION AS A RESULT OF CONTACT WITH SKIN AND OR EYES 7 NOTE orientation of UV lamp pins before REMOVING screws 3 and plates 4 from both ends of the chamber 7 8 REMOVE sea...

Страница 28: ...emove lid from break tank and RELEASE any residual system pressure by PRESSING down the float valve 4 LIFT wing 1 and rotate anti clockwise 5 PIVOT latch 2 to release retaining arm 3 6 SWING retaining...

Страница 29: ...and outlet tubing from the cartridge 4 REMOVE cartridge from clip and discard 5 REMOVE new E cartridge from packaging 6 REMOVE sealing plugs from inlet and outlet ports 7 REMOVE caps from cartridge si...

Страница 30: ...ND IN THE MANUFACTURERS MATERIAL SAFETY DATA SHEET AND ANY APPLICABLE OSHA STANDARDS FOR THE CHEMICAL BEING USED WARNING LABEL THE MACHINE WITH APPROPRIATE WARNING SIGNS SUCH AS DO NOT USE CONTAINS ST...

Страница 31: ...age 29 Sanitization selection screen 3 PRESS to enter sanitization process Enter RO sanitization 4 PRESS to enter RO sanitization Large white lid removal 5 UNSCREW lid on the left of the unit 6 MEASUR...

Страница 32: ...hemical has been added RO sanitization process steps 9 The system now enters a period of automatic sanitization lasting approximately 1 hour Once the step within the sanitization is complete the unit...

Страница 33: ...itization 1 ISOLATE application and ensure the process is OFF 2 PRESENT GREEN sanitization key to reader Sanitization selection screen 3 PRESS to enter sanitization process Enter RO sanitization 4 PRE...

Страница 34: ...to allow the system to empty fill the reservoir The system will stop when the level reaches approximately 10 liters Small white lid removal 8 UNSCREW lid on the right of the unit 9 MEASURE out 250ml...

Страница 35: ...he unit will go straight into process and continue to fill the recirculation reservoir Set sanitization reminder 13 REFIT Labpure L1 LC187 14 PRESS to accept the sanitization reminder The system is no...

Страница 36: ...IN THE MANUFACTURERS MATERIAL SAFETY DATA SHEET AND ANY APPLICABLE OSHA STANDARDS FOR THE CHEMICAL BEING USED WARNING LABEL THE MACHINE WITH APPROPRIATE WARNING SIGNS SUCH AS DO NOT USE CONTAINS STERI...

Страница 37: ...anitization CAUTION ENSURE THE CORRECT DOSAGE OF CHEMICAL IS USED Large white lid removal 5 UNSCREW the large white lid on the left of the unit 6 CUT one EfferSan tablet into four quarters using table...

Страница 38: ...1 hour Once the step within the sanitization is complete the unit will stop and show date re set Set sanitization reminder 11 PRESS to accept the sanitization reminder the unit will go into process a...

Страница 39: ...l CENTRA R 60 120 US MANU40596 Version 02 01 16 Page 37 Sanitization selection screen 3 PRESS to enter sanitization process Enter RO sanitization 4 PRESS to show recirculation sanitization icon 5 PRES...

Страница 40: ...ir The system will stop when the level reaches approximately 10 liters CAUTION ENSURE THE CORRECT DOSAGE OF CHEMICAL IS USED Small white lid removal 8 UNSCREW small lid on the right of the unit 9 CUT...

Страница 41: ...he unit will go straight into process and continue to fill the recirculation reservoir Set sanitization reminder 14 REFIT Labpure L1 LC187 15 PRESS to accept the sanitization reminder The system is no...

Страница 42: ...op over temperature PRESS mute to silence alarm Water temperature is unacceptably high 50 C Contact Technical support 672 93 Protek incorrectly installed PRESS mute to silence alarm CONFIRM Protek L1...

Страница 43: ...ht Add chemical to break tank Option OFF Add chemical to reservoir Option ON Auto restart Output Bell Recirculate Calibration point Reset Cancel Sanitization key Clock Sanitization reminder Data Save...

Страница 44: ...CL2 Silica 30 ppm SiO2 Fouling Index 10 Fl Iron Manganese 0 1 ppm Fe Mn Organics 3 ppm TOC TEMPERATURE 4 40 C Recommended 15 25 C FLOWRATE maximum requirement at 15 C 9 l min Drain Requirements gravi...

Страница 45: ...re alarm Water purity alarm Leak detection alarm Access restricted by User Recognition key Low voltage control circuit 24 V dc Visual alarms Audible alarms Output to operate 24 V dc relay PRODUCT WATE...

Страница 46: ...RTICULAR PURPOSE The warranty provided herein and the data specifications and descriptions of the VWS UK Ltd systems appearing in VWS UK Ltd published catalogues and product literature may not be alte...

Страница 47: ...ELGA LabWater Lane End Industrial Park High Wycombe Bucks HP14 3BY UK Tel 44 0 203 567 7300 Fax 44 0 203 567 7205 E mail techsupport elgalabwater com For the address of your nearest ELGA LabWater Sale...