Page 51
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ELR 9000 HP Series
3.4.4
Adjustment limits
Adjustment limits are only effective on the related set values, no matter if using manual adjust
-
ment or remote control!
Defaults are, that all set values (U, I, P, R) are adjustable from 0 to
102% of the rated value.
This may be obstructive in some cases, especially for protection of
applications against overcurrent. Therefore upper and lower limits
for current (I) and voltage (U) can be set which limit the range of the
adjustable set values.
For power (P) and resistance (R) only upper value limits can be set.
► How to configure the adjustment limits:
1.
On the main screen, tap
to access the SETTINGS menu.
2.
Tap the arrows
to select “
3. Limits”.
3.
In each case a pair of upper and lower limits for U/I or the upper limit for P/R are assigned to the rotary
knobs and can be adjusted. Tap the selection area for another choice
.
4.
Accept the settings with
.
The set values can be entered directly using the ten-key pad. This appears when the touch area
for direct input is tapped (bottom center)
The adjustment limits are coupled to the set values. It means, that the upper limit may not be set
lower than the corresponding set value. Example: If you wish to set the limit for the power set
value (P-max) to 3000 W while the currently adjusted power set value is 3000 W, then the set
value first would have to be reduced to 3000 W or less, in order to set P-max down to 3000 W.
3.4.5
Changing the operating mode
In general, the manual operation of an ELR 9000 HP distinguishes between three operating modes (U/I, P/I, R/I)
which are tied to set value input using the rotary knobs or ten-key pad: This assignment must be changed if one
of the four set values is to be adjusted which is currently not available.
►
How to change the operating mode (two options)
1.
Unless the device is in remote control or the panel is locked, you can switch the
operation anytime. Tap on the depiction of the left-hand knob (see figure to the right)
to change its assignment between U, P and R, or
2.
You directly tap on the coloured areas with the set values, like shown in the figure
to the right. The unit next to the set values, when inverted, indicates the assign-
ment of the knobs. In the example to the right they have P and I assigned, which
means P/I mode.
Depending on the selection the left rotary knob will be assigned different setting values,
the right knob is always assigned to the current.
In order to avoid constant changing of the assignments it is possible, e.g with selection R/I, to
change the other values U and P by direct input. Also see section 3.4.6.
The actual operating mode of the load, which is only effective and indicated while the DC input is switched on,
solely depends on the set values and the input voltage. For more information see section