5. INSTALATION
5.1. Electrical safety and installation.
5.1.1. This equipment is designed for safe use under the following conditions:-
•
Indoor use.
•
Altitude up to 2000 meters.
•
Temperatures between 5
°
C and 40
°
C.
•
Maximum relative humidity 80% for temperatures up to 31
°
C decreasing
linearly to 50% relative humidity at 40
°
C.
•
Mains supply voltage fluctuations up to
±
10% of the nominal voltage.
•
Transient over voltages typically present on the mains supply.(Overvoltage
category II).
•
Applicable rated pollution degree 2.
5.2.
This equipment must be earthed / grounded to a fixed earth / grounded mains
socket outlet. The mains supply is to be earthed / grounded in accordance with
current legislation.
5.3. Ensure only the correct rated mains input fuses are fitted. (Where applicable ensure
the correct Mains cord and moulded IEC plug and lead set fuse if fitted). See
Technical Specification Section 8 of this Instructions book.
5.4. Check the voltage on the product data label on this product unit. Ensure the rating
conforms to your local supply.
5.5. It is recommended this product be connected to a mains supply source which
incorporates an RCD or GFCI device.
5.6.
Do not install this product or accessories on a surface which may
become wet.
5.7. The unit is supplied with a Mains cord wired as follows:
Green / Yellow
or
Green
=
Earth / Ground
Blue
or
White
=
Neutral
Brown
or
Black
=
Live / line hot.
Содержание MH8517