User guide
ELI
O
S
ECH001XN111-A4 – 07/2022
139
Maintenance and servicing
7.3 Malfuction
7.2.2
Accessories
In order to ensure perfect hygiene, it is essential to systematically clean all material and equipment in direct contact
with the patient.
7.3
Malfuction
If you notice a malfunction that is not commented on in the documents accompanying the appliance (see below),
please inform your dealer or the manufacturer.
7.3.1
Possible malfunction
Description of the anomaly
Possible causes
Actions
The device does not start
Discharged battery
Leave the appliance connected to the power
supply for a few hours and then switch it on
again.
Battery Out of order
Contact your distributor to initiate the after-
sales service procedure.
The "Measure" button is not
accessible on the homepage
Memory card out of order
Contact your distributor to change the
memory card
Sound problem during meas-
urement
Check that the acoustic stimu-
lator is correctly connected
Connecting the stimulator
Stimulator is out of order
Contact your distributor to initiate the after-
sales service procedure.
Gas and/or liquid leakage
from the case (during opera-
tion or not)
Battery Out of order
If liquid is leaking or a smell is emitted
from the device even though it is function-
ing properly, it must be returned for ser-
vice. Please contact your dealer to initiate
the service procedure.
Problem transferring data to
PC
Discharged battery
Leave the device connected to the power
supply for a few hours and then resume the
transfer procedure.
- If the transfer still does not work, please
contact your
Distributor.
All consumables (surface electrodes and ear tips) are disposable, discard them after use.
The references of the consumables compatible with your device are listed in the subsection
Erreur !
Source du renvoi introuvable.
.
You can order the consumables from your distributor or directly from
our online shop at www.echodia-store.com
If the device is dropped or if water penetrates, it is imperative to have the device checked by Électronique
du Mazet to exclude any risk (patient and user) related to the use of the device.