Report Level and Temperature Continuously (Factory Diagnostics)
Uuu??
Response: Same as above, except continuously, with internal module configuration and
reported individual switch activation and groups of activated switches.
Note: The unit must be powered down before it will respond to other commands.
Report 4-20mA Output Level (Version 3.09 and Higher)
Uuu?M
Response: UuuMhhhhEeeeeWwwww
uu = unit number
hhhh = hex value 0x0000-0xFFFF eeee = error number
0 = No errors
1= No float detected
2 = One float is out of range on a two-float sensor
3 = Too many groups
4 = (reserved)
5 = Transmit to slave processor for level failed
6 = Transmit to slave processor for temperature failed
7 = Receive from slave processor of level failed
8 = Receive from slave processor of temperature failed
9 = No slave processors responding
Note: If errors 5 through 9 occur and persist after power cycling, the sensor should be
returned for repair.
wwww = warning number
0 = No warnings
1 = Possible level degradation
2 = Possible level degradation due to level offset
Report Temperature Only
Uuu?T
Response: UuuFxxEeeeeWwwww
xx= temperature
Configuration Commands
Assign Unit Number
UuuNnn
uu = unit number (from 00 to 31)
nn = new unit number
Note: Unit number 00 is not valid in Modbus RTU mode
Response: UuuNOK
uu = newly assigned unit number
Содержание DLS 2100 Series
Страница 4: ... 2020 Electrolab Inc March 2020 ...
Страница 31: ... 2020 Electrolab Inc March 2020 RU Flex 2100 DLS Reference Illustrations Figure 1 Model 210 0DLS RU Flex w HLS ...
Страница 32: ... 2020 Electrolab Inc March 2020 Figure 2 Model 2100 DLS RU Flex w HLS Assembly Diagram ...
Страница 34: ... 2020 Electrolab Inc March 2020 Figure 4 Control Drawing 2191 001 ...