© 2020, Electrolab, Inc.
March 2020
Digital Level Sensor Protocol
Command syntax
Uppercase characters denote literals in the command and response streams.
Lowercase characters represent data fields in the command and response
streams. Further explanation of data field structure is provided as necessary
with each command.
All commands are terminated with carriage return <cr>.
All responses are terminated with Ccccc (cccc=16 bit CRC field in hexadecimal)
followed by a carriage return linefeed pair <cr><lf>. All alpha hexadecimal
characters are lowercase.
The prefix to all commands and responses is Uuu where uu is the unit number (00-
31). The unit number is the identity of the level sensor to which a command is
addressed or which generates the response. ‘*’ may be used as a wild card character
for either digit in the unit number field ‘uu’. The responding level sensor will always
convert wildcard characters to the actual unit number.
Commands, which modify a level sensor configuration, always return the
command string and ‘OK’ if successful. ‘EEerr’ replaces ‘OK’ if there is a
problem storing the configuration data in the level sensor EEPROM.
Data Request Commands
Report Level and Temperature
Uuu?
uu is a two-digit unit number from 00 to 31( “*” may be used as a wildcard for either digit)
Note: Do Not use the wildcard “**” if connected to more than one level sensor, as all sensors
will respond simultaneously.
Response: UuuDlll.llFtttEeeeeWwwww
uu = unit number
lll.ll = level in inches (repeated for sensors with 2 floats)
ttt = temperature in degrees F
eeee = error number
0 = No errors
1= No float detected
2 = One float is out of range on a two-float sensor
3 = Too many groups
4 = (reserved)
5 = Transmit to slave processor for level failed
6 = Transmit to slave processor for temperature failed
7 = Receive from slave processor of level failed
8 = Receive from slave processor of temperature failed
9 = No slave processors responding
Содержание DLS 2100 Series
Страница 4: ... 2020 Electrolab Inc March 2020 ...
Страница 31: ... 2020 Electrolab Inc March 2020 RU Flex 2100 DLS Reference Illustrations Figure 1 Model 210 0DLS RU Flex w HLS ...
Страница 32: ... 2020 Electrolab Inc March 2020 Figure 2 Model 2100 DLS RU Flex w HLS Assembly Diagram ...
Страница 34: ... 2020 Electrolab Inc March 2020 Figure 4 Control Drawing 2191 001 ...