MAINTENANCE MAY ONLY BE
CARRIED OUT BY THE RESPONSIBLE
BODY. IT MAY NOT BE CARRIED OUT
BY THE OPERATOR. (THE OPERATOR
MAY ALSO BE THE RESPONSIBILITY
BODY.)
Refer to the instructions for use to know how to use
Tohm-e.
No part of this document may be reproduced. This document consists of a front and back page. © 2015 Electro PJP. All rights
reserved. This issue supersedes all previous issues. Data subject to change without notice.
Notice_TE-DE100_UKMaint. 12/11/2015. Indice A.
Date
Comment
CALIBRATION
AND ADJUSTMENT HISTORY :
SAFETY INFORMATION AND
SPECIFICATIONS :
Protection is compromised if the instructions are not followed.
Device protection :
400 V~ according to EN / IEC 61557-3:2007. IK07
according to EN / IEC 62262.
Operator protection (reference to the earth) :
300 V~ CAT III, reinforced
insulation, class 2, pollution degree 2, according to EN / IEC 61010-1:2010. IP2X
according to EN / IEC 60529.
Operator :
person operating equipment for its intended purpose.
Responsible body :
individual or group responsible for the safe use and
maintenance of equipment.
Environmental conditions :
pollution degree 2 (normal environment) ;
storage and operating temperature range, from -20 °C to +49 °C ; maximum
relative humidity 80 % for temperatures up to 31 °C decreasing linearly to 50 %
relative humidity at 40 °C ; don't submerge the device ; protect if from liquids,
rain, and any precipitation ; do not use it in wet or explosive atmospheres.
Power socket test.
The 'Mains Power On' indicator LED shows that there is
hazardous voltage in reference to the earth, even if the impedance, voltage and
power socket indications are absent. In some cases where power sockets are
incorrectly connected, Tohm-e does not directly indicate the anomaly :
Tohm-e deliberately trips a 30 mA RCD. These are cases where there is no
other way to indicate the fault, e.g. a power outlet with reversed neutral and
PE.
Tohm-e displays earth connection impedance equal to 0 Ω. These are cases,
like case 5 opposite, where the phase is on two contacts at the same time or
the neutral is on two contacts at the same time.
Tohm-e remains fully switched off if the phase is present but if the two other
contacts of the power outlet are not connected. Tohm-e is not a voltage
detector ; do not use it for that purpose.
If Tohm-e remains off even though it is connected to a power socket, check
its
working, for example by connecting it to a power socket known to be in
working condition, before starting work on the power socket.
Power supply
:
power supply from tested power socket (no cell,
accumulator
or battery).
Compliance with standards
EN / IEC61010-1:2010, EN / IEC61557-1:2007,
EN / IEC61557-3:2007, EN / IEC62262, EN / IEC60529, EN61326-1:2013 (basic
electromagnetic environment), EN61326 -2-2:2013, EN61000-3-
2:2006+A1/2009+A2/2009, EN61000-3-3:2008. Compatible with standard NF C
15 -100.
Compliance with European Directives
2011/65/EU 'RoHS', 2006/95/EC
'LVD', 2006/96/EC 'WEEE', 2004/108/EC 'EMC'.
Inspect Tohm-e before use and do not use if damaged.
SOCKET AND EARTH LOOP TESTER
MAINTENANCE
INSTRUCTIONS
Tohm-e TE-DE100. For « German » power sockets, 2P+E, TT.
Compatible with
30 mA RCD.
Pivot head.
For power
sockets, 2P+E,
TT, Germany.
Ergonomically
designed.