- 4 -
www.elechico.com
8.
Setting description
No.
Mode code
Mode Name
Operation
1
A
Check
ambient
temperature
Turn it on, press “+” & “—” button for 3
seconds after the backlight off till the
display shows “A” and value, to check
current ambient temperature.
2
b
Check
blackbody
temperature
Press “+” button once again in “A” mode till
display shows “b” and value, to check
blackbody temperature
3
U
Check battery
voltage
(not
available)
Press “+” button once again in “b” mode till
display shows “U” and value, to check
battery voltage. (not available)
4
F0
ambient
temperature
compensation
Press “+” button once again in “b” mode till
display shows “F0” and value, to check
ambient temp compensation. In this state,
press MODE button into ambient temp
compensation setting, then press “+” or
“—” to set compensation value; when it is
complete, press SET button to confirm it.
5
F1
Body
temperature
compensation
Press “+” button once again in “F0” mode
till display shows “F1” and value, to check
body temp compensation. In this state,
press SET button into body temp
compensation setting, then press “+” or
“—” to set compensation value; when it is
complete, press SET button to confirm it.
6
F2
Blackbody
temperature
compensation
Press “+” button once again in “F1” mode
till display shows “F2” and value, to check
blackbody temp compensation. In this
state, press SET button into blackbody temp
compensation setting, then press “+” or
“—” to set compensation value; when it is
complete, press SET button to confirm it.