R
44
16 PROBE MEASUREMENT PERFORMANCE
Scale
Total range
Accuracy
a
a.
Whichever is the greater. Lower value achieved when calibrated close to the thickness to be measured.
Resolution
in range
F1
F1 2
(F1 mode)
FNF1
N1, N1A
0 µm to 1500 µm
±1% to ±3% or ±2.5 µm
0.1 µm
1.0 µm
0 µm to 99.9 µm
100 µm to 1500 µm
0 mil to 60 mils
±1% to ±3% or ±0.1 mil
0.01 mil
0.1 mil
0 mil to 4.99 mils
5 mils to 60 mils
F1 2
(F2 mode)
N2
0 mm to 5.0 mm
±1% to ±3% or ±0.02 mm
1.0 µm
10 µm
0 mm to 0.99 mm
1.0 mm to 5.0 mm
0 mil to 200 mils
±1% to ±3% or ±1 mil
0.1 mil
1 mil
0 mil to 49.9 mils
50 mils to 200 mils
F3
0 mm to 13 mm
±1% to ±3% or ±0.05 mm
1.0 µm
10 µm
0 mm to 1.99 mm
2 mm to 13 mm
0 mil to 500 mils
±1% to ±3% or ±2.0 mils
0.1 mil
1 mil
0 mil to 99.9 mils
100 mils to 500 mils
FM3
NM3
0 µm to 500 µm
±1% to ±3% or ±2.5 µm
0.1 µm
1.0 µm
0 µm to 99.9 µm
100 µm to 500 µm
0 mil to 10 mils
±1% to ±3% or ±1.0 mil
0.01 mil
0.1 mil
0 mil to 3.99 mils
4 mils to 10 mils
F6
0 mm to 25 mm
±1% to ±3% or ±0.1 mm
10 µm
100 µm
0 mm to 1.99 mm
2 mm to 25 mm
0 mil to 980 mils
±1% to ±3% or ±2.0 mils
1 mil
10 mil
0 mil to 99.9 mils
100 mils to 980 mils
N6
0 mm to 30 mm
±1% to ±3% or ±0.05 mm
10 µm
100 µm
0 mm to 1.99 mm
2 mm to 30 mm
0 mil to 1200 mils
±1% to ±3% or ±2.0 mils
1 mil
10 mil
0 mil to 99.9 mils
100 mils to 1200 mils
Op_456_3_English.book Page 44 Thursday, January 22, 2009 7:58 PM