eTurbotouch kit window
Area to calibration check with guide lines and sample line which meets the criteria for
the correctness of the calibration.
4.1.4. Calibration of single touch panel
Run calibration if touch control works wrong – cursor does not follow pointer position, or if it is necessary to
change orientation of Windows XP (WEPOS Ready 2009) desktop.
1. Run eTurbotouch kit utility and click on „4 pts calibration“ on „General“ marker. Calibration starts.
2. Put finger or pointer nib the most accurate to the center of blinking spot and hold while blinking spot appears on
the other place.
Note: Note that touch point is center of all subjects touch the panel at the time as well as tummy finger center.
Note the perspective – look on the screen upright.
3. Repeat for all calibration points (blinking spots) and press „Save“ after calibration.
4. Check calibration according to previous chapters (testing). If 4-points calibration not work accurate enough,
execute 9 or 25-points calibration. Starting buttons of those calibrations are on „Advance“ marker:
4.1.5. Testing of multi touch panel
The application MosArt RMTP On-line Utility v2.5 is commonly used for the testing. The start icon is situated on
the UniqPC desktop.
1. Run calibration utility MosArt RMTP On-line Utility v2.5.