29
3. MAKEREADY
Fig. 3.14:
Signal table
3.12 STATE past switch ON HV/IS
The device has a mains switch. After the application of the supply voltage,
the device is placed in operation by switching on the mains switch.
After the mains switch is switched on, the built-in mains indicator H1 illu-
minates. The green indicator on the front panel is also briefly illuminated.
If the optional warning lamp set (chapter 7.2) is connected, its green in-
dicator also illuminates.
Devices with touchscreen display
: After booting the operating system
the touchscreen menu appears, see chapter 3.10.
Remote control mode:
A high signal is applied to pin 4 on X7. The device
is controlled using the digital interface (X7) in conjunction with the RS232
/Ethernet/USB interface.
Interface only operation:
A low signal is applied to pin 4 on X7. The
device is controlled only via the RS232 or Ethernet interface.
Two starts mode:
Two start signals are always necessary to apply the
high voltage. These can be: two start commands via the interface, or in
the „Remote mode“ a „digital“ start and a start command over the inter-
face (pay attention to sequence for commands). In the „Local mode“ (front
panel) the green Set button must be pressed twice.
Front panel only
DA interface and RS232/Ethernet/
USB
RS232/Ethernet/USB
only
Operating
mode
Front panel
Digital X7
RS232 X8
Ethernet X9
Only RS232 X8
Ethernet X9/
SK1
SK2
1)
Remote Set button
Remote Set HV
Set PE
Remote
X7-21 X7-15
X7-4
X7-4
X7-2
X7-5
X7-4
off
off
low
low
low
low
low
HV ready for
test
on
on
low
1. Start
high
high (edge) low
low
set HV (1)
HV ON
on
on
low
2. Start
low
set HV (2)
low
set HV (2)
1)
See also Chapt. 6.2.3 "Digital Interface X7"
Note:
Define the value for the test voltage and the Imax limit, as well as
start/stop voltage and ramp speeds and any other parameters as neces-
sary via the touch-screen display or interface (remote mode).
For more detailed information see chapter 10.
Full-electronic G1-1 HV DC-Tester
Содержание G1-1
Страница 38: ...38 4 OPERATING MODES 4 2 2 Normal mode input step Full electronic G1 1 HV DC Tester...
Страница 39: ...39 4 OPERATING MODES 4 2 3 Normal mode output step Full electronic G1 1 HV DC Tester...
Страница 40: ...40 4 OPERATING MODES 4 2 4 Normal mode time step Full electronic G1 1 HV DC Tester...
Страница 133: ...Expires 04 19 2016...