Lactate Scout 

Instruction manual

Содержание Lactate Scout 4

Страница 1: ...Lactate Scout 4 Instruction manual...

Страница 2: ......

Страница 3: ...written consent of SensLab GmbH Part No 7023 9014 0256 Version 1 1 11 2018 Lactate Scout 4 instruction manual Manufacturer SensLab GmbH Bautzner Str 67 04347 Leipzig Germany Sales and service EKF diag...

Страница 4: ...on and off 8 1 5 Display 8 1 6 Code setting for the sensors 9 2 Performing a test 10 2 1 Preparing for a test 10 2 2 Blood collection and test 11 2 3 Checking the code setting 13 2 4 Checking the ambi...

Страница 5: ...splaying stored values 27 3 6 Battery level indicator 28 4 Performing a step test 29 4 1 Step test mode 29 4 2 Heart rate during step tests 31 4 3 Displaying the post exertion period 31 5 Data transfe...

Страница 6: ...ensors Lactate Scout sensors are designed for single use Safety instructions PLEASE READ THE INSTRUCTION MANUAL BEFORE USE AND KEEP IT IN A SAFE PLACE Failure to comply with any text marked as a WARNI...

Страница 7: ...ries that have been provided or recommended by the manufacturer The measuring device must not be used if it is not functioning correctly or has been damaged Your guarantee will be voided and all liabi...

Страница 8: ...etween menus sub menus and menu items for adjusting values OK button 1D for enabling menus sub menus and menu items for confirming choices for activating and deactivating configuration mode Back butto...

Страница 9: ...en inserting the new batteries ensure that the polarity is correct Push the battery cover back in until it completely locks into place 4 If the time to replace the batteries exceeds 30 seconds the dat...

Страница 10: ...r two minutes in configuration and display mode or if the device is not used for two minutes while the sensor is inserted in the measuring device 1 5 Display An information line will be displayed in b...

Страница 11: ...or two seconds 6 Then press the OK button again The code setting menu will open The symbol line will display the Code symbol 7 Enter the two digits of the code starting from the left Enter the digits...

Страница 12: ...up 3 The display will briefly show the code which is currently configured A blinking drop symbol will then be displayed to indicate that the measuring device is ready to take a measurement The ambient...

Страница 13: ...ction and test Wash your fingers or earlobes with clean water at the puncture site You can use the spray bottle supplied with the accessories for this purpose Dry the puncture site In order to take a...

Страница 14: ...A circular symbol with a progress bar will be displayed 5 After ten seconds a second acoustic signal will sound and the test result will be displayed in the measurement unit mmol L The allocated memo...

Страница 15: ...displayed briefly when you switch measuring mode on 7 If more than 25 tests are performed with the same code a reminder to check the code will be displayed for three seconds 8 Pressing the OK button w...

Страница 16: ...vious status 3 Settings functions and displays Settings can be changed in configuration and display mode In order to go to configuration and display mode there must not be a sensor inserted in the mea...

Страница 17: ...ntrol for acoustic signals 2D Device information 2E 1B 2 2A 2B 2C 2D 2E 2D 3A 3B 3C 3 3 1 1 Volume control If you choose the Loudspeaker symbol 2D in the Device settings menu and confirm your selectio...

Страница 18: ...using one of the two arrow keys and confirm your selection by pressing the OK button Once you have confirmed your selection you will then have the option to set the time A clock will appear in the sy...

Страница 19: ...17 2A 4 5 6 7 8...

Страница 20: ...ion by pressing the OK button The sub menu will then open 9 To activate the Bluetooth function use one of the two arrow keys to select the Activation symbol check box with vertical stroke and confirm...

Страница 21: ...by pressing the OK button The measuring device will now search for all Bluetooth Low Energy LE compatible heart rate monitors in the immediate vicinity 12 If the search is successful a list of all of...

Страница 22: ...until the next time that it is activated If you want to delete a registered heart rate monitor simply select the Heart symbol again in the Device settings menu 2B Confirm your selection by pressing t...

Страница 23: ...2 Heart rate If the measuring device is connected to a heart rate monitor the heart rate monitoring system automatically records the measured rate and saves it at five second intervals together with t...

Страница 24: ...s to connect to the heart rate monitor after three automated attempts 20 21 If an individual lactate test is performed while the heart rate is being recorded or a lactate test is performed during a st...

Страница 25: ...used to check the correct functioning of the Lactate Scout 4 measuring system If there are any doubts regarding the accuracy of the test result or the correct functioning of the measuring device you...

Страница 26: ...t mmol L in the symbol line The configured code will be displayed briefly after you have inserted the sensor This code must also match the code printed on the sensor container for the functionality te...

Страница 27: ...track the progress of the test on the display 26 The measured lactate concentration will be displayed immediately afterwards If the result is within the permissible tolerance range a tick will be show...

Страница 28: ...atch function will automatically stop and the measuring device will switch back to the main menu 29 30 1E If you want to stop the stopwatch during the 10 minute interval you can do so by briefly press...

Страница 29: ...e and time 31C of this measurement will be shown in the bottom section of the display If the measurement was taken as part of a step or endurance test and the heart rate was also recorded where applic...

Страница 30: ...isplay options available Batteries are full 33A Batteries are partially discharged 33B Batteries are almost fully discharged the Bluetooth function has been automatically deactivated The batteries nee...

Страница 31: ...cedure for test preparation sampling and measurement is the same as the individual test Insert the sensor into the measuring device You are then in measuring mode Now use one of the two arrow keys to...

Страница 32: ...indicates the number assigned to the measurement exertion level 4 6 Lactate levels are measured at each exertion level during the exertion step test phase The most recently selected step test phase w...

Страница 33: ...e minute leading up to the lactate measurement For this reason it is important to measure the lactate level immediately after each exertion level 4 3 Displaying the post exertion phase During the post...

Страница 34: ...o retrieve the transferred data on your PC you must have the Lactate Scout Assistant software installed This software is available for Windows and Android Additional information is available at www la...

Страница 35: ...aper towel To do this moisten the cloth with a mild cleaning agent You can use water with a small amount of washing up liquid for example Clean the measuring device before wiping it dry WARNING Used m...

Страница 36: ...filling the sensor Blood droplet too small or has run contact time of sensor with blood droplet too short interrupted sample absorption or sensor pressed against skin Repeat the test with a new sensor...

Страница 37: ...sor Check the code configuration Follow the instructions for performing the test Dry your hands well before collecting the blood Use the test solution to perform a func tionality test Please contact t...

Страница 38: ...functionality test with new materials after checking the code Follow the instructions for performing the functionality test Make sure that the target concentration you have selected in the menu match...

Страница 39: ...isplay Time displayed with black background The date and time were not set when the device was configured reset Set the date and time Device fails to turn on Device fails to turn on Device fails to tu...

Страница 40: ...the ambient temperature is within the operating range of the measuring device Change the batteries Please contact the Lactate Scout 4 service team Defective test results Values seem too high too low P...

Страница 41: ...tion Slow update of the screen content Slow update of the screen content The Lactate Scout 4 uses an e pa per display Please note that it may take some time for the display to update depending on the...

Страница 42: ...ct range 35 50 0 5 6 7 mmol L blood lactate 0 2 mmol L 6 8 25 0 mmol L blood lactate 3 Hct ranges 20 35 50 70 0 5 7 5 mmol L blood lactate 0 3 mmol L 7 6 25 0 mmol L blood lactate 4 Measuring principl...

Страница 43: ...480 GHz Transmission power 10 mW SensLab GmbH hereby declares that the Bluetooth radio system type conforms with Directive 2014 53 EU The full EU declaration of conformity text can be obtained via the...

Страница 44: ...itro diagnostic medical device Product meets the requirements of the applicable Directives Manufacturer Serial number Attention consult accompanying documentation observe safety instructions Biologica...

Страница 45: ...er Do not expose to direct solar radiation Protect against liquid Wireless data transfer with Bluetooth wireless technology Device includes RF transmitter Direct current Used batteries must not be dis...

Страница 46: ...r 7023 3405 1008 Lactate Scout test solution Pack of 5 x 2 5 ml bottles with 1 8 2 2 mmol L lactate 7023 6151 0257 Lactate Scout test solution CombiPack of 2 5 ml bottles 2 x 5 units with 4 5 5 6 mmol...

Страница 47: ...0195 PC Pack consists of Dongle for PC for data transfer between Lactate Scout 4 and PC via Bluetooth Low Energy functionality USB stick with Lactate Scout Assistant software for evaluation of lactat...