Notice for this monitor
11
room to a warm room. In that case, do not turn the product on. Instead wait until the dew
condensation disappears, otherwise it may cause some damage to the product.
To Use the Monitor for a Long Time
Quality control
• The display quality of monitors is affected by the quality level of input signals and the
degradation of the product. Perform visual checks and periodic constancy tests
(including grayscale check) to comply with medical standards / guidelines according to
your application, and carry out calibration as necessary. Use of the RadiCS monitor
quality control software enables you to perform high-level quality control that meets
medical standards / guidelines.
• It takes about 15 minutes (under our measurement conditions) for the monitor display to
stabilize. Please wait 15 minutes or more after the monitor power has been turned on or
the monitor has recovered from the power saving mode before performing various tests
for quality control, calibration, or screen adjustment of the monitor.
• We recommend that monitors be set to the recommended level or lower to reduce
changes in luminosity caused by long-term use and maintain stable brightness.
• To adjust measurement results of the integrated calibration sensor (Integrated Front
Sensor) to those of an EIZO external sensor (UX2 sensor) that is sold separately,
perform correlation between the Integrated Front Sensor and the external sensor using
RadiCS / RadiCS LE. Periodical correlation allows you to maintain the measurement
result of the Integrated Front Sensor at a level equivalent to that of the external sensor.
Attention
• The display status of the monitor may change unexpectedly due to an operating error or
unexpected setting change. Using the monitor with the operation switches locked is
recommended after adjusting the screen of the monitor.
For details on how to set, refer to the Installation Manual (on the CD-ROM).
Cleaning
• Periodic cleaning is recommended to keep the product looking new and to prolong its
operation lifetime.
• Stains on the product can be removed by moistening part of a soft cloth with water or by
using our ScreenCleaner and gently wiping.
Attention
• Do not let liquids come into direct contact with the product. If it does, wipe it off immediately.
• Do not allow liquids to get into gaps or inside the product.
• When using chemicals for cleaning or disinfection, chemicals such as alcohol and disinfectant
may cause gloss variation, tarnishing, and fading of the product, and also quality deterioration of
the displayed image. Do not use chemicals on a frequent basis.
• Never use a thinner, benzene, wax, or abrasive cleaner as they may damage the product.
• For more information on cleaning and disinfection, please refer to our web site.
How to check: Access