Appendix
28
EMC Information
The RadiForce series has a performance that appropriately displays images.
Environments of Intended Use
The RadiForce series is intended to be used in Professional healthcare facility environments
such as clinics and hospitals. The following environments are not suitable for the RadiForce
series to be used:
• Home healthcare environments
• In the vicinity of high-frequency surgical equipments such as electrosurgical knives
• In the vicinity of short-wave therapy equipment
• RF shielded room of the medical equipment systems for MRI
• In shielded location Special environments
• Installed in vehicles including ambulances
• Other special environments
WARNING
• The RadiForce series requires special precautions regarding EMC and need to be installed. You
need to carefully read EMC Information and the "PRECAUTIONS" section in this document, and
observe the following instructions when installing and operating the product.
WARNING
• The RadiForce series should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the equipment or system should be observed to verify
normal operation in the configuration in which it will be used.
WARNING
• When using a portable RF communication equipment, keep it 30 cm (12 inches) or more away
from any part, including cables, of the RadiForce series. Otherwise, degradation of the
performance of this equipment could result.
WARNING
• If you are connecting additional equipment to the signal input or signal output parts of the
product and using it as part of a medical system, you must ensure compliance with the
requirements of IEC / EN60601-1-2.