26
Chapter6Troubleshooting
●
Error Code Table
Error Code
Description
0***
• ErrorsthatoccurredduringSelfCalibration
1***
• ErrorsthatoccurredduringGrayscaleCheck
*1**
• ErrorsthatoccurredduringDICOM
*2**
• ErrorsthatoccurredduringCAL1
*3**
• ErrorsthatoccurredduringCAL2
**10
• Theproductmaximumbrightnessmaybelowerthanthetargetbrightness.
• Lowerthetargetbrightness.
**11
• Theproductminimumbrightnessmaybehigherthanthetargetbrightness.
• Raisethetargetbrightness.
**34
• Thesensormaynothaveworkedcorrectly.
• Turnoffthemainpower,waitforafewminutesbeforeturningthepowerbackon,and
thenexecuteSelfCalibration/GrayscaleCheckagain.
**94
• SelfCalibrationorGrayscaleCheckcannotbeexecuted.
• EntertheNativeresolution.
**95
• TheexecutionconditionsforSelfCalibrationorGrayscaleCheckmaynotbe
appropriate.