8
About the monitor
About the monitor
Intended Use
This product is intended to be used to display medical images, such as endoscopic surgery images.
Attention
• This product is not intended for diagnostic purposes.
• This product should be set to horizontal view mode.
• This product may not be covered by warranty for uses other than those described in this manual.
•
The specifications stipulated in this manual are only applicable when the enclosed power cord is used.
•
Only use optional products manufactured or specified by us with this product.
Precautions for Use
• Parts (such as the LCD display, fan, etc.) may deteriorate from long-term usage. Periodically check that
they are operating normally.
• When the screen image is changed after displaying the same image for extended periods of time, an
afterimage may appear. Use the screen saver or power save function to avoid displaying the same image
for extended periods of time.
• If the monitor displays continuously over a long period of time, dark smudges or burn-in may appear. To
maximize the life of the monitor, we recommend the monitor be turned off periodically.
• An afterimage may appear even after a short time period when a blinking image is displayed in the same
position, because of LCD panel characteristics. If this occurs, displaying a white image or a moving
image may solve the issue.
•
The backlight of the LCD panel has a fixed lifetime. When the screen becomes dark or begins to flicker,
please contact your local EIZO representative.
• The screen may have defective pixels or a small number of light dots on the screen. This is due to the
characteristics of the panel itself, and is not a malfunction of the product.
• Do not press on the panel or edge of the frame strongly, as this may result in display malfunctions, such
as interference patterns, etc. If pressure is continually applied to the panel, it may deteriorate or damage
it. (If the pressure marks remain on the panel, leave the monitor with a black or white screen. The
symptom may disappear.)
• Do not scratch or press on the panel with any sharp objects, as this may result in damage to the panel.
Do not attempt to wipe with tissues as this may scratch the panel.
• Condensation may form on the surface or interior of this product when it is brought into a cold room,
when the temperature suddenly rises, or when it is moved from a cold room to a warm room. Also, if
the air conditioner is turned on after this product has been kept in a warm room or a room with high
humidity for an extended period, change the facing or placement of this product so that the air from the
air conditioner does not blow directly on this product. If the air from an air conditioner blows directly on
to the display surface, condensation may form on the inside of the protection panel. In that case, do not
turn the monitor on. Instead wait until the dew condensation disappears, otherwise it may cause some
damage to the monitor.
• It takes about 30 minutes for the monitor display to stabilize. Please wait 30 minutes or more after the
monitor power has been turned on or the monitor has recovered from the power saving mode, and then
adjust the monitor.