30
Chapter4SelfCalibration
●
Setting the calibration schedule “Schedule”
Set the calibration schedule for SelfCalibration.
Attention
• Clock adjustment is necessary for SelfCalibration to operate.
Procedure
1. Choose “SelfCalibration” from the Adjustment menu, and press
.
2. Choose “Settings” from “SelfCalibration”, and press
.
3. Choose “Schedule” from “Settings”, and press
.
The schedule setting menu appears.
4. Use or to select the items to change, and press
.
Setting Item
Setting
StartingTime
Off
SelfCalibrationisnotexecuted.
Setthetimingatwhich
SelfCalibrationisexecuted.
PowerSave
Whenthetimeperiodsetin
theschedulehaselapsed,itis
executedifthemodeinitially
switchestoPowerSave,orifthe
powerinitiallyswitchesoff.
Immediately
Whenthetimeperiodsetin
theschedulehaselapsed,itis
executedimmediately.
SettingType
*
Type1
Onceeveryfewmonths
Setaccordingtothefrequencyat
whichSelfCalibrationisexecuted.
Type2
Onceeveryfewweeks
Type3
Onceeveryfewhours(monitor
usagetime)
Interval
Type1
1-12months
Settheintervalbetweenthe
completedSelfCalibrationandthe
nextexecution.Thesettingdetails
differdependingonthetypeof
“SettingType”specified.
Type2
1-5weeks
Type3
50-500hours
Week
1stweek-5thweek
Setwhichweektoexecute
SelfCalibrationonforthemonthin
whichcalibrationisexecuted.
Day
SUN/MON/TUE/WED/THU/FRI/SAT
Setwhichdaytoexecute
SelfCalibrationonfortheweekin
whichcalibrationisexecuted.
Time
00:00to23:30
SetthetimeatwhichSelfCalibration
isexecuted.
*
Settingsthatareusedfrequentlycanbepreset.
5. Use or for adjustment.
6. Press
to exit.
The calibration schedule is set.
Note
• When the starting time is “Immediately”, and the setting type is either “Type1” or “Type2”, SelfCalibration is
executed according to schedule even if the power is switched off using .
• When the starting time is “Off”, the SelfCalibration menu shows “Setting incomplete.” However, SelfCalibration can
be executed manually.