EIT UV Power Puck II Скачать руководство пользователя страница 1

 EIT Part Number P/N IM-0111 Rev A Issued April 2019  

 

 
 
 
 

 

 

 
 

EIT

®

 

UV Power Puck

®

 II & UviCure

®

 Plus II 

User’s Guide 

 

 
 
 
 

 

 

Electronic Instrumentation and Technology (EIT) LLC 
309 Kelly’s Ford Plaza SE 
Leesburg VA 20175 USA  
Phone: 703-478-0700 
Email: 

[email protected]

 

Web: www.eit.com 

 

 

Содержание UV Power Puck II

Страница 1: ...1 Rev A Issued April 2019 EIT UV Power Puck II UviCure Plus II User s Guide Electronic Instrumentation and Technology EIT LLC 309 Kelly s Ford Plaza SE Leesburg VA 20175 USA Phone 703 478 0700 Email u...

Страница 2: ...tics Error Messages 13 Low Battery Indicator 13 Over Temperature State 13 Over Range State 13 Other Error Codes 13 Maintenance Cleaning and Calibration 14 Cleaning Display 14 Calibration 14 EIT UV Ins...

Страница 3: ...The instrument reading includes total energy and peak irradiance of all four transmission bands for the UV Power Puck II and one band for the UviCure Plus II The EIT UV Power Puck II and UviCure Plus...

Страница 4: ...ns are shown below Setup Mode Button On Off Button Depress and Hold until screen is illuminated Run Exit Run Button Press to enter Run Mode Press to Exit Run Mode Display Battery Door Unit Serial Numb...

Страница 5: ...er Puck II Internal Firmware Version Serial Number Last Calibration Date Dynamic Range 10 Watts and UV bands UVA UVB UVC UVV Unit Type UviCure Plus II Internal Firmware Version Serial Number Last Cali...

Страница 6: ...A short press of the RUN button clears the memory and puts the unit in the RUN mode The display shows RUNNING after shortly displaying the internal temperature of the unit Confirm that the unit displa...

Страница 7: ...tons located under the arrows to scroll in the indicated direction To change the default selection first select the line then the setting on each line Press the SAVE button to save the setting as the...

Страница 8: ...ine The data will remain until it is overwritten The difference or change between the current run data and the reference data is displayed as a percentage change on the DIFF line TOGGLE UviCure Plus I...

Страница 9: ...ng an instrument outside of the Suggested Operating Ranges can lead to variations in the readings o If the unit is used well below the Suggested Operating Range there can be variations in the Joules e...

Страница 10: ...n the display each time RUN mode is activated An alarm will sound when the unit reaches an internal temperature of 65 C or greater Avoid pre heat flash off and or IR sections of your line Repeated slo...

Страница 11: ...uld lead to variations especially in the irradiance values As technology improved the sample rates in our instruments have also dramatically increased The PowerMAP when released had a user adjustable...

Страница 12: ...normally referred to as the Average Peak Intensity or Smooth Profiler intensity This example shows a 0 30 second section of the file collected at 2048 Hz It is clear from this example that the power s...

Страница 13: ...sampled and is reported Smooth Smoothing The UviCure Plus II and Power Puck II oversample at a very high sample rate The user is able to adjust the effective sample rate to one of three settings in th...

Страница 14: ...y Power Puck and UviCure Plus units ends December 31 2019 SMOOTH PROFILER SMOOTH PROFILER displays the Peak Irradiance at an effective sample rate of 128 samples per second 128 Hz This rate matches th...

Страница 15: ...and can be read by pressing the Select button When doing this the beeping tone stops and you can scroll through the data readings In addition if the internal temperature of the unit exceeds 75o C the...

Страница 16: ...ns provided in Appendix A or on our website https www eit com products care and cleaning Calibration EIT s calibration period for the Power Puck II or UviCure Plus II is six months This based on our h...

Страница 17: ...You will be responsible for damage incurred from inadequate packaging if the original packaging is not used The customer is responsible for insuring the unit during transportation to EIT Please see o...

Страница 18: ...t alteration negligence improper installation lack of routine maintenance or other causes beyond our control 3 The goods have been repaired altered or modified by anyone other than EIT or EIT authoriz...

Страница 19: ...ating on optics window also most often cause the readings to increase 5 There are two cleaning methods Wipes Swabs described on this Guide Select the one that best suits your needs and train your staf...

Страница 20: ...more away from the optics Short bursts from a distance will minimize the transfer of any additives from the canned air to the optics If using compressed air make sure it is oil free instrument grade a...

Страница 21: ...nges UVICURE Plus II One channel monitoring of UVA 320 390 nm UVB 280 320nm UVC 250 260nm or UVV 395 445nm selected at the time of purchase Spatial Response Approximately cosine Lambertian Operating T...

Страница 22: ...hanges in speed and power It also allows the user to break down the contribution from each individual lamp on multi lamp systems EIT s UV PowerView III Software for Profiler enabled instruments is ava...

Страница 23: ...e Plus II that shipped prior to January 1 2017 included a basic software program that transferred only the Joule and Watt numerical values to a computer screen This basic software program which worked...

Страница 24: ...Class B device in accordance with Council Directive 98 34 EEC Technical Standards Regulations Directive procedure for the provision of information in the field of technical regulations and rules on in...