Displays weight with enhanced resolution for 3 seconds.
The “
Dd/10
”
symbol
above the key will flash when enhanced resolution is shown.
Switch between Gross and Net reading.
Not used.
T
Sets the net weight to zero and select Net reading.
>0<
Sets the gross weight to zero and select Gross reading.
–
Perform automatic dosing
To perform an automatic dosing, follow this procedure:
1.
The desired Coarse and Fine setpoints must be entered. Use the
key in the
NORMAL
screen to select the
DOSING
screen. Then use the
DOSING
screen to enter the desired
setpoints.
2.
To start a new dosing
select the “
Action: START DOSING
” using the
or
key in the
DOSING
screen and then press the
key. Alternately start a new dosing by activating
the digital
START
input.
3.
This will zero the net weight (automatic tare) and activate the digital
COARSE
and
FINE
dosing outputs.
4.
During dosing the
COARSE
output is active as long as the net weight does not exceed the
COARSE limit. An active
COARSE
output indicates dosing should be done at high speed
for the first part of the dosing, followed by slow speed for the final part of the dosing
where the
COARSE
output is deactivated.
5.
When the net weight (positive or negative) reaches the FINE setpoint (possibly adjusted
by the afterflow correction) the
FINE
output will be deactivated and dosing is stopped.
6.
Hereafter the
COARSE
and
FINE
outputs will remain deactivated until a new dosing is
started.
7.
The dosed amount is automatically registered after the entered registration period, if
the registration period is different from 0 ms. In this case a registration can be made by
activating the
REGISTRATION
input. The result of the last registration can be read in the
NORMAL
screen.
8.
Following a registration the totals are updated and a possible new afterflow correction
may be calculated.
9.
An ongoing dosing can be stopped prior to
reaching the setpoint, by selecting the “
Ac-
tion: STOP DOSING
” using the
or
key in the
DOSING
screen and then pressing the
key. An ongoing dosing is also aborted if the weighing range is exceeded or if a load
cell error occurs.
10.
The weighing terminal is ready for start of a new dosing.