1.3 A short exercise
Let’s do a quick exercise to show how easy the Scuba is to use. We will connect the instrument to a PC,
calibrate conductivity, and check the calibration log for the conductivity calibration information.
1
Connect your Scuba to your PC with the USB adapter and either a data cable or underwater cable.
2
Click the Scuba icon to launch the Scuba control software. Soon you will see the home page and the data
being sent from your Scuba.
3
Fill the storage/calibration cup with tap water or conductivity standard and wait for the conductivity
reading to stabilize.
4
Click the Scuba pull-down menu on the home page and click
Calibrate
, and then click
SpCond
on the next
screen.
5
Type in the approximate conductivity of your tap water or the value of your conductivity standard and
click
OK
.
6
Click on
OK
in the next screen to finalize the calibration and be returned to the home page. (D.5)
7
Click the Scuba pull-down menu on the home page then click
Cal Log
on the next screen. At the bottom of
the list are the details of the calibration that you just did.
8
Click on
OK
to return to the home page.
Was that easy, or what? Don’t you wish you had back all those frustrating, life-shortening, soul-crushing hours
you spent trying to decipher those other multiprobe manufacturers’ software?