background image

 

27 

There are several conventional ways to calibrate fluorometers; those are explained in Calibrating Eijkelkamp’s Turner 

Fluorometers. Eijkelkamp also has a new method for “standardizing” fluorometers if you do not use a conventional 

calibration method; that is explained in Standardizing Eijkelkamp’s Turner Fluorometers. Both of these documents can be 

found on the flash drive. 

 

The maintenance procedure is pretty much the same for all fluorometers: 

1

 

Cleaning the sensor: Rinse the chlorophyll sensor in fresh water following each deployment, ideally until it is 
completely clean again. Do not let the chlorophyll sensor come in contact with any organic solvents, such as acetone 
and methanol, or strong acids and bases. 

2

 

Cleaning the optics: Visually inspect the optical window after each deployment following a soaking in fresh water. 
Use optical tissue to clean the window with soapy water, if needed. 

4.15 Ion-Selective Electrodes (ISE’s) 

 

ISE’s are traditionally used in the laboratory at a constant, moderate 

temperature, with ionic strength adjusters added to each sample so that 

the sample and calibration solution have roughly the same ionic 

strength. ISE’s can provide valuable information in the field, for instance 

in watching short-term trends, but their calibration requirements, 

stability, and accuracy are not nearly as simple and reliable as those for 

DO, conductivity, turbidity, etc. For field use, Eijkelkamp makes several 

theoretical and empirical corrections to elicit the best possible field 

performance. 

 

ISE’s operates much like a pH electrode except that the pH glass is replaced by a membrane that is selective for the 

analyte of interest (ammonium, chloride, nitrate, sodium). The electrode’s filling solution contains a salt of the analyte, 

and the difference between that salt’s concentration and the analyte concentration in your water produces a charge 

separation. That charge separation is measured, relative to the reference electrode, as a voltage that changes predictably 

with changes in the analyte concentration in the water adjacent the membrane. 

 

It’s best not to let your ISE dry out, so place a small amount of tap water in the storage cup to ensure 100% humidity. The 

sensing elements (tip) for nitrate and ammonium ISE’s have lifetimes of about 90 days. Then, you must replace the tip by 

unscrewing it from the sensor body and screwing in a new tip. 

Each sensor body is programmed for a specific ISE; the ISE is identified by a series of rings or dots. One ring or dot means 

the sensor body is programmed for a chloride ISE; two mean calcium, three mean nitrate, four mean ammonium, 5 mean 

bromide, and six mean sodium.  

 

  

It’s a good idea to limit the submersion of ISE’s to about 10 meters. If you need to go deeper than that with the  
other sensors, you can use a plug for the ISE electrode

.

 

 

Note that the ammonium ISE senses ammonium, but at pH’s higher than about 8 the ammonium (NH4+) is mostly 

converted into ammonia gas (NH

3

). Eijkelkamp’s software uses the pH, conductivity, and temperature of the sample 

water to calculate ammonia (as mg/L-N). You can also display total ammonia; the sum of ammonia and ammonium. 

Note that ammonium and nitrate ISE’s suffer interference from positive ions, especially potassium and sodium, and 

sodium ISE’s suffer interference from positive ions, especially potassium and ammonium. The chloride ISE does not 

normally suffer from interfering ions. 

 

Note that Eijkelkamp’s sodium ISE has a plastic membrane with a wider pH range (pH 3 – 10) and less pH interference 

than the traditional sodium ISE’s (which are made with glass membranes). This sensor is specified to have a 10-second 

response time and a range of 0.05 – 2,300 mg/L Na+. The sodium ISE can be immersed to 15 meters of water without 

damage, but there may be a pressure effect on the reading. Although Eijkelkamp testing on a limited sample size showed 

this error to be within our ±20% accuracy specification, we recommend that users check the performance of their 

particular sensors under actual field conditions. 

Note that that the sodium ISE has a slow response to changes in temperature and may take many minutes to reach a final 

reading when the temperature changes significantly. 

If your Scuba is equipped with more than one ISE, use care when replacing tips so that you don’t put a tip on the wrong 

sensor (for example put a nitrate on the sodium sensor). 

 

Содержание Scuba

Страница 1: ...3 88 02 00 E info eijkelkamp com I royaleijkelkamp com 2022 07 M Scuba Scuba and Scuba Trimeter water quality probes This manual covers the Royal Eijkelkamp water quality probes Scuba and Scuba Trimet...

Страница 2: ...ete a custom parameter 18 3 6 Software updates 18 4 Sensors and calibrations 19 4 1 The difference between sensors and parameters 19 4 2 Basics of parameter calibrations 19 4 3 Choosing calibration st...

Страница 3: ...ns ScubaLink App 44 7 Frequently asked questions 45 7 1 Our three most frequently asked questions 45 7 2 Sensors 45 7 3 Calibration and maintenance 47 7 4 Communication and software 49 7 5 Deployment...

Страница 4: ...one 3 The storage calibration cup protects the sensors when the Scuba is not in use keep a few ounces of tap water in the storage calibration cup when the Scuba is not being used And with the lid remo...

Страница 5: ...w opens click Install Scuba software to upload the Scuba user interface software and the USB driver software onto your PC Depending on which version of Windows you are accursed with you might have to...

Страница 6: ...ur Scuba to the USB adapter using either a Scuba data cable or underwater cable Click the Scuba icon to connect your PC and Scuba The home page will appear showing the Scuba s real time data and vario...

Страница 7: ...onductivity reading to stabilize 4 Click the Scuba pull down menu on the home page and click Calibrate and then click SpCond on the next screen 5 Type in the approximate conductivity of your tap water...

Страница 8: ...eks at a time with cable supplied power or an optional Scuba integral battery pack You can for instance set the instrument to take a set of readings every half hour anchor it in an estuary and return...

Страница 9: ...ed on the Amphibian 3 mirrors the version for the PC with some concessions for the small screen Section 5 has more information about small screen data displays 2 6 What do the LED lights mean The Scub...

Страница 10: ...els etc 2 9 Flow cell If it is more convenient to bring the water to the Scuba than the Scuba to the water for instance when monitoring a ground water well you can simply screw a flow cell onto the Sc...

Страница 11: ...11 3 The Scuba control software 3 1 Scuba control software map...

Страница 12: ...s are These two buttons are pull down menus These eight buttons are hot buttons short cuts to important and often used functions This area is continuously updating real time data from your Scuba The b...

Страница 13: ...data to PC with annotation hot button to save the most recent line of data as shown in the yellow band on the home page to your snapshot file along with a brief note that you might wish to append to...

Страница 14: ...e active snapshot file location is listed on the bottom line of the home page 3 Click Automatic Snapshot if you wish log data automatically and quickly for instance to catch a transient situation or i...

Страница 15: ...n your Scuba Highlight the file you re interested in by clicking on it Then with the other buttons on the screen you can then view that file on your data display screen delete that file or export it t...

Страница 16: ...he check mark and disables that parameter Note that if you enable a parameter but don t have a sensor for that parameter it would be a huge coincidence if the data were accurate The order of the enabl...

Страница 17: ...viant SRF and you will be returned to the home page If you click on Cancel you ll go back to the calibrate screen How do I know if I need to calibrate The simple answer is that frequent calibration wi...

Страница 18: ...is the same measurement as Turb NTU except altered slightly to make its true ISO 2027 response linear with formazin 2 The Sensors and Parameters List now contains the parameter Turb_Mod which is the...

Страница 19: ...ater especially if you have been using other calibration solutions Pour the water into the calibration cup position the stopper side of the lid on top and shake the Scuba vigorously to remove traces o...

Страница 20: ...pH reference 2 or 3 points pH 4 pH 7 pH 10 pH 7 pH 10 most common ORP 1 point ORP Standard 200 mV Conductivity 1 point CD Standard 0 5 Molar 58670 Micro S CD Standard 0 1 Molar 12856 Micro S CD Standa...

Страница 21: ...80 When you press the OK button to accept a calibration the Scuba automatically accepts your calibration if the SRF is between 60 and 140 If the SRF falls outside that range you will be cautioned to...

Страница 22: ...manufacturers of optical DO sensors recommend that you not calibrate the zero DO point However we support zero DO calibration in the Scuba software and think it s a good idea to check your sensor s z...

Страница 23: ...te the voltage measuring circuit The pH glass is specially formulated to absorb water so that ions particularly H and OH in the water are attracted to the glass to offset the ionic constituency of the...

Страница 24: ...e measuring circuit Because platinum does not react with ions in the water it won t give or take any electrons from those ions unless they are very persuasive The potential voltage created by this ref...

Страница 25: ...at 90 to that beam More particles in the water mean more of that light is scattered so the turbidity reading is higher Any material that accumulates on the optical surfaces of the turbidity sensor is...

Страница 26: ...ody and sliding a fresh brush into the brush slot as shown below Use the first Allen wrench from your wiper kit to remove the standard turbidity sensor wiper Use the other Allen wrench to install the...

Страница 27: ...st not to let your ISE dry out so place a small amount of tap water in the storage cup to ensure 100 humidity The sensing elements tip for nitrate and ammonium ISE s have lifetimes of about 90 days Th...

Страница 28: ...ry out the membrane between stations by using only a few drops of water in the storage cup instead of a few ounces 4 17 PAR The PAR Photosynthetically Active Radiation sensor measures the amount of li...

Страница 29: ...tric Pressure BP if you have a Clark or optical dissolved oxygen sensor so click on the Set Barometric Pressure button it s in the Calibrate menu You can set the BP by typing the correct value in mm H...

Страница 30: ...the Scuba logging time uniformity For instance if your logging interval is 15 minutes and you turn logging on at five minutes past 10 AM your first data will be logged at exactly 10 15 and then every...

Страница 31: ...ght ways to power a Scuba 5 7 Logging with an Internal Battery Pack All Scubas may be ordered with an optional Internal Battery Pack IBP a watertight housing with a cassette for batteries that is perm...

Страница 32: ...isture dirt grit and any other debris off the exposed O ring surfaces and the inside of the battery sleeve Add a small amount of silicone grease to the O rings and to the inside of the battery sleeve...

Страница 33: ...5 8 When you are ready to deploy the Scuba turn the battery switch to ON And turn the switch back to OFF when you retrieve the Scuba 9 Remember to look for the red LED to blink five times to confirm t...

Страница 34: ...n all moisture dirt grit and any other debris off the exposed O ring surfaces and the inside of the battery sleeve Add a small amount of silicone grease to the O rings and to the inside of the battery...

Страница 35: ...s six pins with a gap in the outer circle of pins 4 Line up the white dot with the gap in the connector pins and slowly push them together Very little force is needed 5 Grasping only the Scuba and EBP...

Страница 36: ...ur specific application We recommend using the highest quality alkaline batteries available such as Duracell Copper Tops Rechargeable batteries can be used but their battery life is typically only hal...

Страница 37: ...he sensors in copper ions that discourage biological growth 2 The extended turbidity brush attaches to the turbidity sensor and cleans the measurement surfaces of several sensors including the Turner...

Страница 38: ...get you through a full day of field work The Bluetooth battery s Bluetooth address is shown on a label on the back side 6 4 Establish Bluetooth communication Follow the directions below for establish...

Страница 39: ...to MORE Smartphone or tablet and select Android Scan Filter Clear any settings in field Select OK Go to the Bluetooth SETTINGS on the data display and select the Bluetooth ID of your Scuba Bluetooth...

Страница 40: ...40 Snapshot Files Create New Snapshot File View Snapshot File Contents Snapshot Annotation Home Page Snapshot Menu...

Страница 41: ...41 Snapshot Report Calibration Conductivity 1 Calibration Conductivity 2 Calibration Acceptance About Amphibian Software Version Calibrate Menu...

Страница 42: ...42 Set Scroll Interval Calibrate Set Time and Date PC Sync Report Calibration Barometric Pressure 1 Calibration Barometric Pressure 2 Sensor and Parameter List...

Страница 43: ...43 Graphs Turned On Manage Scuba Files Logging Setup Interval and Wiping Logging Setup Browse Logging Setup Interval and Wiping Scuba Scuba log Scuba...

Страница 44: ...44 6 5 2 Screens ScubaLink App...

Страница 45: ...ical DO caps that never need replacement 8 turbidity mimicking software 9 PDA based data display 10 Bluetooth connectivity 11 Cell phone as a data display 12 a digital instruction manual 13 smart sens...

Страница 46: ...replace DO caps haven t you It s OK lots of people have been in the same boat Now that you are a Scuba owner you can expect your DO caps to last five years or more You can see the condition of your c...

Страница 47: ...n with a high range turbidity standard 4 3 4 What is an SRF Suppose that a typical conductivity sensor reports 100 A in a 1413 S cm standard If your particular conductivity sensor reports 100 A in tha...

Страница 48: ...about 1000 S cm then calibrate with the readily available 1413 S cm KCl standard Note that some sensors not SC have two calibration points the second point should be set at a convenient low point usua...

Страница 49: ...st people should never have to worry about choosing a COM port just let the PC do the work 3 4 5 Is the coloured top line an average of the values or the latest readings No the data in the coloured ba...

Страница 50: ...l components regardless of how many sensors you order Most importantly we have a no cramming rule that prevents our stuffing too many sensors into an artificially small instrument diameter Yes we know...

Отзывы: