12 EMC
Tables
12.EMC Tables
Guidance and manufacturer’s declaration – electromagnetic emissions
The
E-connect S
is intended for use in the electromagnetic environment specified below.
The customer or the user of the
E-connect S
should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The
E-connect S
uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The
E-connect S
is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions
IEC61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The
E-connect S
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
E-connect S
should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV
air
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV
air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30 %.