Cleaning and Instruction Manual 

EickView 50/ 60/ 70 

Videoendoscope 

 

Содержание EickView 50

Страница 1: ...Cleaning and Instruction Manual EickView 50 60 70 Videoendoscope ...

Страница 2: ...fety 2 2 Applicable Scope 4 3 The notice of use 4 4 Configuration 4 5 Operation 6 6 Cleaning and Sterilization 10 7 Storage 11 8 Functional Check 12 9 Malfuction and Solution 13 10 Technical Specifications 14 ...

Страница 3: ... and attention label Signal Meaning Warning Warning Warning indicates in case of neglect of attention the presence of a hazard that could result in severe personal injury death or substantial property damage Note Note indicates informs the user in case of neglect of attention to be improper installation operation or maintenance such behavior can lead to device damage 1 2 Environment factors 1 2 1 ...

Страница 4: ...t the authorized person Please immediately shut off the power when the device damage appears abnormal image or can t bend after cleaning and disinfection This equipment may only be used together with accessories supplied by manufacturer Please put the manual at the place where the operator and maintainer can easily get Before every operation visually check the equipment for exterior mechanical dam...

Страница 5: ...anual Please immediately clean disinfect dry and store the videoscope in a designated location after using this equipment The package of this device can not be pressed impacted and knocked so as to prevent deformation damage 4 Configuration 4 1 This equipment consists of several components portable VET videoscope 3 5 LCD display power adaptor AV output cable USB cable SD card leakage tester vent c...

Страница 6: ...Insertion part Operation part It s the holding and operation part Leakage tester Connect to the adaptor for leakage tester to check the seal of videoscope Vent Cover During gas sterilization transportation or long term storage ventilation cover should be installed on the adaptor for the leakage tester while in the cleaning liquid immersion disinfection it should be removed 3 5 LCD Supply the power...

Страница 7: ...ots on the lock of LCD and the adaptor of videoscope on a same level connect the LCD on the adaptor of videoscope and lock clockwise tightly The LCD can be rotated 5 2 Insert SD card Lift the protective cover on the left side of the LCD face up SD card into the slot completely then close the over If there is no card inside of LCD please do not to take photo or video If you want to take the SD card...

Страница 8: ...ng channel Open the sealing part insert the biopsy forceps into instrument channel in the center of the rubber cover Biopsy Forceps Rubber cover of adaptor of instruments Note Do not overexert when bending If the bending resistance is too much or it can not in place or cannot bend it should be returned to service repair When the insertion part set up do not touch the pole of bending to avoid damag...

Страница 9: ...ing part Note Do not insert the instruments when the distal part is bending to avoid damage the inner parts 5 6 Recording Press the camera button on the LCD or button1 on the operation part to take photos Press the video button on the LCD or button 2 to take the video To browse the photos or videos please choose menu and check the files The details of operation of 3 5 inch display please refers to...

Страница 10: ...lease the button Button 3 5 8 Disassemble Turn off the power of 3 5 LCD after using switch off the adaptor of LCD and take it off from the videoscope 5 9 Charge You need to charge when low battery of the LCD It takes about 10 hours to make fully charge The red light will be off if the charging is finished Pull out the power plug at this time ...

Страница 11: ...If necessary clean the SD card with alcohol cotton Disinfected solution must be clearly indicated suitable for endoscopes Note Before cleaning ensure the sealing of videoscope Do not install the vent cover After using the videoscope please clean it immediately Do not use any cloth or tools which can damage any surfaces of the device Cleaning Brush There are two methods of disinfection for the port...

Страница 12: ... of air pressure For liquid disinfection the vent cover must be removed Otherwise it would damage videoscope 7 Storage Storage condition see 1 2 Environment factors During the long term storage the vent cover should be covered and the insertion part and bending section should be placed straightly Charge the battery of 3 5 inch display every three months Note For long term storage the insertion par...

Страница 13: ...ot use the videoscope again and contact service After leakage testing take out the videoscope from the solution Screw off the deflated knob to the finger of presser meter to zero Take off the leakage tester from videoscpe It would be damaged without deflation The adaptor of leakage tester The deflated knob Note Before each operation please do leakage testing Please do not use the videoscope if the...

Страница 14: ...ir Check if the button on the 3 5 LCD is in failure or not If it does not work properly please contact our company for repair 9 3 Bending failure The bending angle is not in place send to the manufacturer or service When bending resistance is too much Do not use it and immediately contact service If it can not be bent Do not use it and immediately contact service 9 4 Frozen image If the 3 5 LCD is...

Страница 15: ...14 ...

Страница 16: ...15 ...