Environmental conditions
The following general requirements must be fulfilled:
•
The room must have exhaust ventilation.
•
The instrument should not be exposed to corrosive gases.
•
The instrument should not be exposed to sources of heat, such as direct sunlight.
•
Dust in the atmosphere should be kept to a minimum.
•
Strong magnetic or electric fields should be avoided.
•
The instrument should not be exposed to vibrations.
The installation site for XDUO 2500 Mixer must comply with the following specifications:
Specification
Parameter
Indoor use only
Allowed location
5°C to 30°C
Ambient temperature, operation
20% to 85% RH, non-condensing
Relative humidity, operation
-25°C to 60°C
Ambient temperature, transportation
-25°C to 50°C
Ambient temperature, storage
Up to 2000 m
Altitude, operation
Electrical power
[
Requirement
Parameter
100-120 V AC, 50 or 60 Hz, 12 A
Supply voltage, XDUO 2500 Mixer
220-240 V AC, 50 or 60 Hz, 6 A
See the manufacturer's documentation.
Supply voltage, TCU
Two outlets for the following units:
Number of electrical outlets
•
XDUO 2500 Mixer
•
TCU
126
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
5 Installation
5.2 Site requirements