![Edwards nEXT200200 Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/edwards/next200200/next200200_instruction-manual_2366136063.webp)
The venting option is now stored in the memory of the pump.
To check the venting option that is set, send a query as follows:
Command
?
S
8
5
3
cr
The reply is as follows:
Reply
=
S
8
5
3
sp
d
cr
5.1.4 Standby speed setting
You can operate the pump at a standby speed rather than full rotational speed. The
standby speed is a user-configurable option and can be set to the value between 55%
and 100% full rotational speed. When the pump is shipped, it is configured with a
standby speed of 70% full rotational speed.
To change the standby speed setting, use one of the methods that follows:
▪
Use the STDBY+ and STDBY- push buttons (given on the controller). Each push of
the button adjusts the speed by 10 Hz (1%) Refer to
▪
Use a serial command
To change the standby speed setting using a serial command, send the command that
follows (characters 'd' shows the value as a percentage of full rotational speed):
Command
!
S
8
5
7
sp
d
d
d
cr
The reply is as follows:
Reply
*
S
8
5
7
sp
r
cr
The standby speed is now stored in memory of the pump.
To check what standby speed is set, send a query as follows:
Command
?
S
8
5
7
cr
The reply is as follows:
Reply
=
S
8
5
7
sp
d
d
d
cr
5.1.5 Normal speed setting
The normal speed is a user-configurable setting and can be set to the value between
50% and 100% full rotational speed. When the pump is shipped, it is configured with a
normal speed of 80% full rotational speed.
To change the normal speed setting, send the command that follows (characters 'd'
shows the value as a percentage of full rotational speed):
Command
!
S
8
5
6
sp
d
d
d
cr
The reply is as follows:
Reply
*
S
8
5
6
sp
r
cr
The normal speed is now stored in the memory of the pump.
Page 59
B85200880_D - Configuration
Содержание nEXT200200
Страница 6: ...This page has been intentionally left blank...
Страница 24: ...Figure 3 nEXT 200200 NW25 interstage port pumping speed Page 20 B85200880_D Technical data...
Страница 77: ...6 8 Bakeout nEXT200200 split flow pumps are not applicable for bakeout Page 73 B85200880_D Operation...
Страница 100: ...edwardsvacuum com...