![Edwards nEXT200200 Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/edwards/next200200/next200200_instruction-manual_2366136029.webp)
If the power limit setting is not adjustable, use a power supply capable to supply
sufficient current to meet manufacturer's factory default power limit setting shown in
Table: Logic interface technical data on page 27
.
The manufacturer's TIC has variants. If the pump is operated using the
manufacturer's TIC, contact us to select the applicable variant for the application.
3.7 Pumping media
WARNING: DANGEROUS GASES
Risk of asphyxiation. Release the dangerous gases and gas mixtures safely. Do not
expose people to the gases. If hazardous gases or vapour must pump, obey the safety
recommendations of the supplier of the gas or vapour.
WARNING: PYROPHORIC OR EXPLOSIVE GASES
Risk of explosion. Do not use the nEXT pump to pump pyrophoric or explosive gas
mixtures as the pump is not applicable for this purpose. The pump and its
connections are not designed to contain an explosion.
WARNING: DANGEROUS GASES
Risk of damage to equipment. In the interstage and booster versions of the
nEXT pumps, gas pumped through the interstage port will mix with gas pumped
through the pump inlet. Make sure that the gases do not react or combine to form
dangerous gases and substances.
WARNING: HIGH OPERATING PRESSURE
Risk of damage to equipment. Do not exceed the maximum continuous operating
pressure. Failure to do so can cause dangerous rotor temperatures and will decrease
the life of the pump.
WARNING: VACUUM EXPOSURE
Risk of injury. Do not expose the part of the human body to the vacuum.
WARNING: MERCURY VAPOUR
Risk of damage to equipment. Do not use a nEXT pump to pump mercury vapour. Do
not allow mercury (for example, from a McLeod gauge) to come in contact with the
pump. If mercury vapour is pumped, the pump rotor can corrode and fail.
Note:
Concentrations of gases can be changed by the compression of the pump.
The pumps are designed to pump the residual gases used in high
‑
vacuum systems are:
▪
Air
Page 25
B85200880_D - Technical data
Содержание nEXT200200
Страница 6: ...This page has been intentionally left blank...
Страница 24: ...Figure 3 nEXT 200200 NW25 interstage port pumping speed Page 20 B85200880_D Technical data...
Страница 77: ...6 8 Bakeout nEXT200200 split flow pumps are not applicable for bakeout Page 73 B85200880_D Operation...
Страница 100: ...edwardsvacuum com...