2.4.5 Controller configuration (serial configuration)
All configuration settings stored in the motor controller are kept when the power to the
pump is removed so that it is possible to use another system to configure the motor
controller before the installation of the pump at the application. You can configure the
operation of the pump as per the application and the pump can be operated using a
simple parallel interface system.
To configure the pump, use a customer simple serial system or use the
manufacturer's TIC Turbo Controller or Turbo Instrument Controller. The TICs allows the
storage of a pump's configuration. You can download the configuration to another
pump. This is useful when you configure a number of pumps with the same settings
before they are installed at a system.
Note:
The RS485 / RS232 slide switch must be in the RS232 (default) position if the TIC is to be
used to configure the pump. Refer to
Connect the serial interface to the customer control
.
The TIC is supplied with a Windows
TM
based PC program which allows the pump to be
configured from a single PC. The program has a simple user interface and it is not
necessary to use the ASCII message protocol given in
Connection for serial control and
. The TIC PC program can save number of pump configurations
which can be downloaded to the other pumps.
Page 16
B85200880_D - Introduction