Figure 28 Serial and parallel control flowchart
6.6 Operation with a TIC or TAG
The pump can be connected to manufacturer's TIC Turbo Instrument Controller, TIC
Turbo Controller or TAG Controller. The TIC gives the power necessary to operate the
pump, but the TAG requires a separate PSU to be connected.
Instructions on the setup and operation with the TIC Turbo Instrument Controller, TIC
Turbo Controller or TAG Controller can be found on CD ROM part number P450-00-000,
which is supplied with the TIC or TAG.
6.7 Decelerating and venting
Decelerate the pump before venting.
The system can be vented in accordance with the advice given in
on page 72.
CAUTION: RATE OF INCREASE IN PRESSURE
Risk of damage to equipment. Do not open a manual vent valve until the pump
rotational speed decreases below 50%, otherwise the rate of increase in pressure can
be too high, which can damage the pump. In an emergency, open the vent valve
quickly to decelerate the pump rotor in the shortest possible time.
If using the motor controller to control a TAV solenoid valve, there is a 2
‑
second delay
between:
07/2021 - ©Edwards Limited
Page 81
B8G000880_E
B8G000880_E - Operation