Remedy
Refer to
Unplanned shut down and alarms
on page 80 to determine the exhaust
pressure that caused the warning.
Check that all valves in the exhaust line are open and consider whether process
debris or condensation are likely.
Fault
Warning 39.13 - Exh Press Fault
Cause
Sensor is faulty or disconnected
Remedy
Contact us.
Fault
Warning 54.11 - MB Mtr Temp High
Cause
The temperature of the booster pump is too high.
Remedy
Check that cooling water is connected, switched on and is to specification given in
on page 21.
Fault
Warning 54.13 - Sensor missing
Cause
The booster temperature sensor has possibly become disconnected or failed.
Remedy
Check that the sensor is installed and is correctly connected.
page 21
Fault
Warning 54.13 - Sensor Fault
Cause
Sensor is faulty (short circuit).
Remedy
Check that the sensor is installed and is correctly connected.
Fault
Warning 55.11 - DP Temp High
Cause
Pump temperature high
Remedy
Check that cooling water is connected, switched on and is to specification given in
on page 21.
Fault
Warning 55.13 - Sensor missing
Cause
The dry pump temperature sensor is possibly disconnected or failed.
Remedy
Check that the sensor is installed and is correctly connected.
Fault
Warning 55.13 - Sensor Fault
Cause
Sensor is faulty (short circuit).
Remedy
Check that the sensor is installed and is correctly connected.
10/2020 - ©Edwards Limited
Page 71
M56635880_D
M56635880_D - Fault finding
Содержание iXM Series
Страница 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Страница 101: ...This page has been intentionally left blank...
Страница 102: ...This page has been intentionally left blank...
Страница 103: ...This page has been intentionally left blank...
Страница 104: ...edwardsvacuum com...