Technical data
B722-40-880 Issue H
Figure 3 - Podule connector plug
Table 9 - Logic interface connector pins
Pin Number Signal
Polarity Use
2
0 V Control reference
-
0 V reference for all control and status signals
3
START/STOP control input
-
Connect to Pin 2 to start pump
4
STANDBY control input /
Serial RX
-
Connect to Pin 2 to enable standby speed
5
Serial enable
-
Connect to Pin 2 to enable serial interface mode
7
FAIL / Serial TX
-
Logic high when fail condition exists in parallel mode
9
Analogue output
Positive 0 - 10 V output proportional to measured output
10
Chassis / Screen
-
Screen
12
Chassis / Screen
-
-
15
NORMAL status output
-
Logic low when pump rotational speed is at normal speed or
above
8, 13, 14
Electrical supply: 0 V
-
1, 6, 11
Electrical supply: 24 V
Positive
Table 10 - Podule technical data
Description
Data
Connector plug
Phoenix part number FKMC1881325
Voltage output
24 V d.c. +10%, -20% (19.2 to 26.4 V d.c.)
Current output
120 mA
1. Negative terminal
2. Positive terminal