Technical data
B756-01-880 Issue K
2
Technical data
2.1
General
2.2
Pumping media
Table 1 - General
Performance
Dimensions
See
and
Maximum inlet flange temperature
70 °C
Maximum magnetic field
3.5 mT Horizontal field
Installation category
EN61010 part 1, Category 1
Pollution degree
EN61010, Pollution degree 2
Electrical supply
24 V d.c.
Equipment type
Fixed equipment, for indoor use only
Ambient operating temperature range
5 ºC to 40 ºC
Ambient operating humidity range
10 to 90% RH (non-condensing)
Maximum operating altitude
2000 m
Ambient storage temperature range
-30 ºC to 70 ºC
WARNING
Vent dangerous gases and gas mixtures safely. Do not expose people to these gases.
WARNING
Do not use Split Flow Pumps to pump explosive gas mixtures as the pumps are not suitable for this
purpose.
WARNING
If pyrophoric gases are pumped then the customer must supply adequate dilution with an inert gas.
WARNING
It is the customer’s responsibility to leak check the system to ensure process gases cannot escape
to the atmosphere and cause a hazard.
WARNING
Gases pumped from different inlets may combine to cause dangerous mixtures.