B756-01-880 Issue K
Page 8
© Edwards Limited 2013. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
Starting time to 90%
speed
EXDC160
110 se
110 sec
120 sec
150 sec
EXC100E
190 sec
190 sec
EXC100L
245 sec
EXC120
130 sec
130 sec
EXC300
100 sec
100 sec
120 sec
EXDC80
270 sec
330 sec
Cooling method
Forced air/
water
Forced air/
water
Ambient air temperature
(forced-air cooling)
0 - 35 °C
0 - 35 °C
0 - 35 °C
0 - 35 °C
Water temperature
(water cooling)
10 - 20 °C
10 - 20 °C
-
-
Noise level (at 1 metre)
< 50 dB(A)
< 50 dB(A)
< 50 dB(A)
< 50 dB(A)
Recommended
controller
EXDC160
EXDC160
EXDC160
EXDC160
EXDC160 maximum VA
input
250 VA
250 VA
250 VA
250 VA
Quiescent power
25 W
25 W
40 W
40 W
Recommended backing
pump
‡‡‡
RV3
RV3
RV3
RV3
*
The data shown for EXT200/200H GCMS applies to pumps: B75640991, B75641991, B75642991, B75643991
and B75644991. This is because these pumps have the same internal components and the data is taken
without inlet screens fitted.
†
Mass applies to products listed under the Declaration of Incorporation.
‡
Pb = backing pressure,
P
i
inlet pressure,
**
Pumping speeds are without inlet screen or inlet strainer (EXT200/200Hi, GCMS and ICPMS). Inlet screens
and inlet strainers reduce speed by approximately 10%.
††
This is a measured value. Theoretical value >10
9
and >10
11
respectively for GCMS and ICPMS.
‡‡
Theoretical values 2 x 10
9
and 7 x 10
11
respectively.
***
Inlet pressure has risen to 10
-3
mbar.
†††
Above this pressure, rotational speed drops below nominal.
‡‡‡
A larger backing pump may be required for maximum throughput.
Table 4 - Technical data (continued)