Introduction
1.9.3
Axial Emergency (EXC300M only)
The EXT250M pump has sensors which detect the axial displacement of the pump rotor; the outputs of these sensors
are monitored by the Controller.
The Controller trips into Axial Emergency condition if:
z
The displacement of the rotor exceeds set limits for more than 2 seconds.
z
The displacement of the rotor repeatedly exceeds set limits (due to oscillation).
When the Controller trips into Axial Emergency condition, the electrical supply to the pump-motor is switched off,
the Emergency LED on the front panel goes on and the Axial Emergency status output signal on the logic interface
opens. The EXT250M pump electromagnetic bearings remain active, although the rotor assembly may contact the
safety bearings if the axial loading exceeds the capabilities of the bearing drive.
1.10
Electrical connections
All electrical connections to the Controller are on the rear panel: see Figures